The need for standardization of tacrolimus assays.

Daniel M Levine1, Gregory T Maine, David A Armbruster, Christopher Mussell, Christoph Buchholz, Gavin O'Connor, Victoria Tuck, Atholl Johnston, David W Holt.   

Abstract

BACKGROUND: Owing to the lack of an internationally recognized tacrolimus reference material and reference method, current LC-MS and immunoassay test methods used to monitor tacrolimus concentrations in whole blood are not standardized. The aim of this study was to assess the need for tacrolimus assay standardization.
METHODS: We sent a blinded 40-member whole-blood tacrolimus proficiency panel (0-30 μg/L) to 22 clinical laboratories in 14 countries to be tested by the following assays: Abbott ARCHITECT (n = 17), LC-MS (n = 9), and Siemens Dade Dimension (n = 5). Selected LC-MS laboratories (n = 4) also received a common calibrator set. We compared test results to a validated LC-MS method. Four samples from the proficiency panel were assigned reference values by using exact-matching isotope-dilution mass spectrometr at LGC.
RESULTS: The range of CVs observed with the tacrolimus proficiency panel was as follows: LC-MS 11.4%-18.7%, ARCHITECT 3.9%-9.5%, and Siemens Dade 5.0%-48.1%. The range of historical within-site QC CVs obtained with the use of 3 control concentrations were as follows: LC-MS low 3.8%-10.7%, medium 2.0%-9.3%, high 2.3%-9.0%; ARCHITECT low 2.5%-9.5%, medium 2.5%-8.6%, high 2.9%-18.6%; and Siemens/Dade Dimension low 8.7%-23.0%, medium 7.6%-13.2%, high 4.4%-10.4%. Assay bias observed between the 4 LC-MS sites was not corrected by implementation of a common calibrator set.
CONCLUSIONS: Tacrolimus assay standardization will be necessary to compare patient results between clinical laboratories. Improved assay accuracy is required to provide optimized drug dosing and consistent care across transplant centers globally.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21998339     DOI: 10.1373/clinchem.2011.172080

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  9 in total

Review 1.  Clinical Evaluation of Modified Release and Immediate Release Tacrolimus Formulations.

Authors:  Simon Tremblay; Rita R Alloway
Journal:  AAPS J       Date:  2017-07-17       Impact factor: 4.009

Review 2.  Clinical Chemistry Laboratory Automation in the 21st Century - Amat Victoria curam (Victory loves careful preparation).

Authors:  David A Armbruster; David R Overcash; Jaime Reyes
Journal:  Clin Biochem Rev       Date:  2014-08

3.  Long-term cross-validation of everolimus therapeutic drug monitoring assays: the Zortracker study.

Authors:  Björn Schniedewind; Stefanie Niederlechner; Jeffrey L Galinkin; Kamisha L Johnson-Davis; Uwe Christians; Eric J Meyer
Journal:  Ther Drug Monit       Date:  2015-06       Impact factor: 3.681

4.  Harmonisation of laboratory testing.

Authors:  Jillian R Tate; Roger Johnson; Kenneth Sikaris
Journal:  Clin Biochem Rev       Date:  2012-11

5.  A guide to harmonisation and standardisation of measurands determined by liquid chromatography - tandem mass spectrometry in routine clinical biochemistry.

Authors:  Ronda F Greaves
Journal:  Clin Biochem Rev       Date:  2012-11

Review 6.  Population Pharmacokinetics of Tacrolimus in Transplant Recipients: What Did We Learn About Sources of Interindividual Variabilities?

Authors:  Olivia Campagne; Donald E Mager; Kathleen M Tornatore
Journal:  J Clin Pharmacol       Date:  2018-10-29       Impact factor: 3.126

7.  Tablet or capsule form of generic mycophenolate mofetil (My-Rept®) after liver transplantation: a prospective randomized trial.

Authors:  Jong Man Kim; Jong Wook Oh; Sangjin Kim; Jinsoo Rhu; Ji Soo Lee; Kyeong Sik Kim; Gyu-Seong Choi; Jae-Won Joh
Journal:  Drug Des Devel Ther       Date:  2019-07-02       Impact factor: 4.162

8.  Immunosuppressive Drug Measurement by Liquid Chromatography Coupled to Tandem Mass Spectrometry: Interlaboratory Comparison in the Korean Clinical Laboratories.

Authors:  Hyun-Ki Kim; Hyung-Doo Park; Sang-Guk Lee; Hyojin Chae; Sang Hoon Song; Yong-Wha Lee; Yeo-Min Yun; Sunhyun Ahn; Serim Kim; Sun Min Lee; Soo-Youn Lee; Sail Chun
Journal:  Ann Lab Med       Date:  2021-05-01       Impact factor: 3.464

9.  Analytical performance evaluation of the Elecsys® Cyclosporine and Elecsys® Tacrolimus assays on the cobas e411 analyzer.

Authors:  Maki Sasano; Shigeki Kimura; Ikuhiro Maeda; Yoh Hidaka
Journal:  Pract Lab Med       Date:  2017-03-10
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.