Electrocardiograms transmitted via facsimile may not allow accurate interval interpretation.

Kanwal M Farooqi1, Scott R Ceresnak, Katherine Freeman, Robert H Pass.   

Abstract

BACKGROUND: Electrocardiograms (ECGs) are sent via facsimile by pediatricians and psychiatrists to cardiologists for assessment. The validity of this method of transmission has not been established.
METHODS: ECGs were collected from 100 consecutive patients from the cardiology clinic of the Children's Hospital at Montefiore. The ECGs were faxed and also electronically faxed (efax) and printed. Two electrophysiologists (EP1 and EP2) interpreted the intervals on original, faxed, and efaxed ECGs and intervals (RR, PR, and QT) were compared. A three-way analysis of variance to examine differences between raters, among ECG intervals, and among methods (repeated factor) was performed. Because no interaction terms were significant, a Duncan's multiple range test was used to evaluate where differences occurred among the three intervals and three methods, given these main effects were significant.
RESULTS: The difference between raters EP1 and EP2 was not significant (P = 0.6681). Although the interval measurements of the faxed and efaxed ECGs were not significantly different from each other (P > 0.05), each was significantly different from the original across all three ECG intervals and both raters (P = 0.0138). The RR interval yielded mean (SD) values for the original, faxed, and efaxed methods of 0.6986 seconds (0.2074), 0.6646 seconds (0.1938), and 0.6838 seconds (0.1935), respectively. For the QT interval, the mean (SD) values for the original, faxed, and efaxed methods were 0.3370 seconds (0.0524), 0.32134 seconds (0.0466), and 0.3284 seconds (0.0515), respectively.
CONCLUSIONS: ECG transmission via facsimile or efax may introduce significant distortion of the intervals. Alternative means of sending ECGs for interpretation should be considered. ©2011, The Authors. Journal compilation ©2011 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21995491     DOI: 10.1111/j.1540-8159.2011.03158.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Faxing ECGs from peripheral hospitals to tertiary paediatric cardiology units--is it safe and sustainable?

Authors:  Sam McGrath; Greg Skinner; Gareth J Morgan
Journal:  Ulster Med J       Date:  2014-01

2.  Improving ECG Services at a Children's Hospital: Implementation of a Digital ECG System.

Authors:  Frank A Osei; Gregory J Gates; Steven J Choi; Daphne T Hsu; Robert H Pass; Scott R Ceresnak
Journal:  Int J Pediatr       Date:  2015-09-14
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.