Literature DB >> 21985092

An inter-laboratory comparison of urinary 3-hydroxypropylmercapturic acid measurement demonstrates good reproducibility between laboratories.

Emmanuel Minet1, Graham Errington, Gerhard Scherer, Kirk Newland, Mehran Sharifi, Brian Bailey, Mike McEwan, Francis Cheung.   

Abstract

BACKGROUND: Biomarkers have been used extensively in clinical studies to assess toxicant exposure in smokers and non-smokers and have recently been used in the evaluation of novel tobacco products. The urinary metabolite 3-HPMA, a metabolite of the major tobacco smoke toxicity contributor acrolein, is one example of a biomarker used to measure exposure to tobacco smoke. A number of laboratories have developed liquid chromatography with tandem mass spectrometry (LC-MS/MS) based methods to measure urinary 3-HPMA; however, it is unclear to what extent the data obtained by these different laboratories are comparable.
FINDINGS: This report describes an inter-laboratory comparison carried out to evaluate the comparability of 3-HPMA measurement between four laboratories. A common set of spiked and authentic smoker and non-smoker urine samples were used. Each laboratory used their in-house LC-MS/MS method and a common internal standard. A comparison of the repeatability ('r'), reproducibility ('R'), and coefficient of variation for 3-HPMA demonstrated that within-laboratory variation was consistently lower than between-laboratory variation. The average inter-laboratory coefficient of variation was 7% for fortified urine samples and 16.2% for authentic urine samples. Together, this represents an inter-laboratory variation of 12.2%.
CONCLUSION: The results from this first inter-laboratory comparison for the measurement of 3-HPMA in urine demonstrate a reasonably good consensus between laboratories. However, some consistent measurement biases were still observed between laboratories, suggesting that additional work may be required to further reduce the inter-laboratory coefficient of variation.

Entities:  

Year:  2011        PMID: 21985092      PMCID: PMC3208382          DOI: 10.1186/1756-0500-4-391

Source DB:  PubMed          Journal:  BMC Res Notes        ISSN: 1756-0500


Background

Cigarette smoke contains thousands of chemicals, including toxicants, which can be categorized as either gases, semi-volatiles (gas/vapor phase), or particles ("tar" phase) [1]. Machine-measured cigarette yields under the ISO testing regimen do not provide an accurate estimate of human exposure to cigarette smoke toxicants [2]. These limitations have led to the development of methods to quantify biomarkers for specific toxicants in biological fluids such as urine, saliva, and plasma [3]. The gas phase, tobacco smoke toxicant acrolein [CAS:107-02-8] (Figure 1A) has been identified by the World Health Organization (WHO) study group on Tobacco Product Regulation (TobReg) as a major contributor to smoke toxicity [4]. This evaluation was based on the concentration of acrolein in smoke and its toxicity potency factor (cancer and non-cancer), established using various models. 3-hydroxypropylmercapturic acid (3-HPMA) is the major urinary metabolite of acrolein (Figure 1B) [5], and it can be quantified using LC-MS based methods [3,5].
Figure 1

Chemical structure of acrolein (. 3-HPMA and 3-HPMA-13C3-15N molecular weights are also indicated.

Chemical structure of acrolein (. 3-HPMA and 3-HPMA-13C3-15N molecular weights are also indicated. One critical element in the measurement and interpretation of biomarker data (including 3-HPMA) is the comparability in method analysis between different laboratories, which can use different methodologies. For instance, a study conducted by Biber and colleagues on a common set of urinary and plasma samples comparing nicotine and cotinine data from eleven laboratories concluded that individual values could vary significantly between laboratories [6]. In a more recent study, Bernert and colleagues showed that good measurement reproducibility for cotinine in a common set of samples could be achieved between six laboratories, when a standardized HPLC-UV method was used [7]. In this study we tested the reproducibility of 3-HPMA measurement between four laboratories using their in-house method and a common set of fortified and authentic urine samples. Each laboratory used a common reference compound and the internal standard 3-HPMA-13C3-15N.

Material and methods

Reagents and samples

Synthetic 3-HPMA (reference compound, Figure 1B) and 3-HPMA-13C3-15N (internal standard, Figure 1C) were obtained from AptoChem (Montreal, Canada). 3-HPMA-13C3-15N was ordered as a custom synthesis and the same lot was used by each laboratory. 3-HPMA-d3 was supplied by Toronto Research Chemicals (North York, Canada). Pooled non-smoker urine samples were supplied fortified with 3-HPMA by RECIPE Chemicals (Munich, Germany). Four concentrations of synthetic 3-HPMA were used: unspiked (background ≈30-50 ng/ml), 400 ng/ml, 1200 ng/ml, and 3600 ng/ml 3-HPMA. Prior to distribution, the samples were quantified using one of the participating laboratories (Lab 2) in order to ensure the quality of the preparation. The samples were then portioned into 5-ml-aliquots, lyophilized and shipped to the laboratories in triplicates of each sample (3 × 4 vials with lyophilized urine). The laboratories were advised to reconstitute the samples with 5 ml water. Five authentic urine samples, covering a 3-HPMA concentration range which reflects typical levels in non-smokers to heavy smokers, were aliquoted in triplicates and sent to the participating laboratories (3 × 5 vials). The lyophilized non-smoker urine samples were supplied by RECIPE® (Munich, Germany), a supplier of samples used for quality assurance testing. The smoker samples were obtained as part of a biomarker study conducted previously by BAT. The corresponding study protocol and informed consent forms were approved by the Ethics Committee of the Bayerische landesarztekammer Munich, Germany (v. 18.02.2008), which contained a provision for revisiting the samples for the purpose of biomarker method development. The clinical study was conducted in accordance with the World Medical Association Declaration of Helsinki (World Medical Association, 2004) and International Conference on Harmonisation (ICH) Guidelines for Good Clinical Practice (GCP) (International Conference on Harmonization, 1996).

Analytical procedure

The samples were labelled with a code number and randomized prior to distribution. 10 mg of each reference compound and the internal standard were dispatched as a dried powder. Samples were received in one batch by each laboratory in September 2009, and all analyses were completed by November 2009. A summary of the analytical methods is given in Table 1, which is based on the performance of QC samples measured prior to the initiation of the study. All four laboratories used their in-house protocol, which included a solid phase extraction step and analysis by LC-ESI-MS/MS. The methods were validated according to FDA guidelines.
Table 1

Analytical method characteristics

Laboratories1234
MethodLC-ESI-MS/MSLC-ESI-MS/MSLC-ESI-MS/MSLC-ESI-MS/MS
modepositivenegativepositivenegative
SPE (column)Phenomenex Strata-XIsolute ENV+Waters OASISWaters OASIS
SPE recovery (%)74687578
HPLC column (make)Waters Xterra MS C18Waters HILIC-SilicaWaters Acquity PhenylThermo BioBasic AX
HPLC column (size)50 × 2.1 mm, 2.5 μm150 × 2.1 mm, 3 μm100 × 2.1 mm, 1.7 μm50 × 3 mm, 5 μm
Quantifier ion (mass)m/z 222 - 163m/z 220 - 89m/z 222 - 117m/z 220 -91
Qualifier ion (mass)m/z 222 - 117m/z 220 - 91
precision intra-day (%)1.1 to 5.91.4 to 8.61.1 to 1.51.2 to 6.5
precision inter-day (%)a5.1 to 5.33.3 to 71.7 to 3.93.3 to 7.5
accuracy (%)93.2 to 10283.9 to 10297.6 to 10296.8 to 101
LOD (ng/ml)2.21NDdNDNDd
LOQ (ng/ml)72550f35f
Linearity (ng/ml)7 to 540025 to 1000050 to 500035 to 5000
Matrix effect (%)ND7.4 to 17-1.7 to 19.6-5.5 to 6.3

aValues represent the precision range obtained for low, medium, and high concentrations at the time the methods were developed, except for Lab 1 where only low and high concentrations were tested.

bBased on calibration standards.

cLOD was an estimate based on spiked water, water being used as the SPE solvent.

dNot determined

eLOQ was established in spiked water, water being used as SPE solvent.

fMatrices used were either diluted non-smoker urine or non-smoker urine with very low 3-HPMA background.

Analytical method characteristics aValues represent the precision range obtained for low, medium, and high concentrations at the time the methods were developed, except for Lab 1 where only low and high concentrations were tested. bBased on calibration standards. cLOD was an estimate based on spiked water, water being used as the SPE solvent. dNot determined eLOQ was established in spiked water, water being used as SPE solvent. fMatrices used were either diluted non-smoker urine or non-smoker urine with very low 3-HPMA background.

Statistical analysis

Basic statistical analysis was carried out with MINITAB v15.1. Individual value plots were produced to test inter-laboratory variation of 3-HPMA concentrations. A non-parametric Wilcoxon paired t-test was performed to compare preliminary test analyses conducted with different internal standards. Analysis of covariance was used to compare the analytical methods at the four laboratories using the same internal standard (3-HPMA-13C3-15N). Precision statistics, as defined in ISO 5725-2 [8], were used as a measure of random errors, and expressed as repeatability ('r') and reproducibility ('R'). For the purposes of this study, in which each laboratory used its own method, 'R' refers to inter-laboratory variation.

Results and discussion

Data for both fortified and authentic urine, using 3-HPMA-13C3-15N as internal standard, were reported by each lab and the corresponding 3-HPMA concentrations (ng/ml) are shown in Table 2. In addition, laboratory 1 repeated the measurements using two different internal standards - 3-HPMA-d3 and 3-HPMA-13C3-15N - which were prepared and analyzed on the same day, in order to investigate the potential confounding effects of using different standards under the same analytical conditions. A non-parametric paired t-test showed that the use of 3-HPMA-d3 gave consistently higher concentrations than 3-HPMA-13C3-15N (Figure 2), highlighting the importance of standardizing the use of internal standards across each laboratory throughout the study.
Table 2

All urinary 3-HPMA data in ng/ml

3-HPMA (ng/ml)

SamplesLab1Lab2Lab3Lab4
1apooled NSa urine40.631.9< LOQb< LOQ
bpooled NS urine29.830.8< LOQ< LOQ
cpooled NS urine40.030.7< LOQ< LOQ

2afortified NS urine466492396405
bfortified NS urine402470400403
cfortified NS urine431471401402

3afortified NS urine1302134011801070
bfortified NS urine1230127011601140
cfortified NS urine1140134011501090

4afortified NS urine3624378034703220
bfortified NS urine3714382034203210
cfortified NS urine3504397033703240

5aNS urine39.848.3< LOQ< LOQ
bNS urine48.246.1< LOQ36.6
cNS urine62.446.9< LOQ35.1

6asmoker urine371376269321
bsmoker urine376382294269
csmoker urine371370300258

7asmoker urine870842613556
bsmoker urine960874673659
csmoker urine840830721619

8asmoker urine10801200862969
bsmoker urine11221160925914
csmoker urine10441180878929

9asmoker urine1482139012001100
bsmoker urine1296137012101130
csmoker urine126013901100894

Samples are numbered from 1 to 9. Samples 1 non-smoker urine, samples 2-4 = fortified samples, and samples 5-9 = authentic urine samples. Each sample was aliquoted in triplicates labeled a, b, and c.

aNS: non smokers

b< LOQ: below limit of quantification

Figure 2

Boxplot of non-parametric differences between 3-HPMA measured in all samples using 3-HPMA-d. The hypothesis (Ho) is based on no difference (0) between the 3-HPMA-d3 measures minus the 3-HPMA-13C3-15N measures. The box plot shows a clear positive difference with p = 0 based on a Wilcoxon paired t-test with a 95% confidence interval for the mean difference (x).

All urinary 3-HPMA data in ng/ml Samples are numbered from 1 to 9. Samples 1 non-smoker urine, samples 2-4 = fortified samples, and samples 5-9 = authentic urine samples. Each sample was aliquoted in triplicates labeled a, b, and c. aNS: non smokers b< LOQ: below limit of quantification Boxplot of non-parametric differences between 3-HPMA measured in all samples using 3-HPMA-d. The hypothesis (Ho) is based on no difference (0) between the 3-HPMA-d3 measures minus the 3-HPMA-13C3-15N measures. The box plot shows a clear positive difference with p = 0 based on a Wilcoxon paired t-test with a 95% confidence interval for the mean difference (x). A background level of 40 to 60 ng/ml 3-HPMA was observed in the non-smoker samples selected for this study. This is expected given that acrolein is also the product of lipid peroxidation, fossil fuel combustion, and is found in cooked food [9]. As a quality control check, data from the fortified urine samples (Table 2) were plotted to generate a regression line and the corresponding equation. Using this, the values for the urine samples (Table 2) were recalculated based on the 3-HPMA peak area and the 3-HPMA-13C3-15N internal standard. The calculated concentrations were consistent with the reported concentrations from each lab (Additional file 1). Individual value plots constructed using the sample data indicate a close similarity in the measurements, across the broad range of 3-HPMA concentrations, for all four laboratories using 3-HPMA-13C3-15N (Figure 3). However, an analysis of covariance (ANOVA) indicated a significant variation between laboratories still existed (Table 3). A closer fit could be observed between lab 1 and 2, and between lab 3 and 4. The coefficients of variation, giving an estimate of the imprecision for repeated measures at different concentration ranges, are also reported in Table 3. The imprecision for each concentration range should be interpreted carefully since the replicate measures were obtained from three aliquots from a single solution rather than a triplicate measure of a unique sample. The fortified samples (samples 2 to 4), were used as an internal calibration reference to calculate accuracies (Table 4).
Figure 3

Individual value plots for 3-HPMA (blue circles) (ng/ml). A. Value plot for the fortified samples in four laboratories. B. Value plot for authentic urine samples in four laboratories. Missing values were below the LOQ.

Table 3

One way ANOVA for 3-HPMA vs laboratories (lab1, 2, 3, and 4) for each set of samples

Mean 3-HPMA (ng/ml)StDevCoVP value (one-way Anova)
Sample 1lab136.86.116.50.186
lab231.130.72.1
lab3< LOQNANA
lab4< LOQNANA

Sample 2lab143332.27.40.002
lab2477.712.42.6
lab33992.60.7
lab4403.31.50.4

Sample 3lab1122481.26.60.004
lab21316.740.43.1
lab31163.315.31.3
lab4110036.13.3

Sample 4lab13614105.42.90.000
lab23856.7100.22.6
lab33420501.5
lab43223.315.30.5

Sample 5lab150.111.422.80.182
lab247.11.12.4
lab3< LOQNANA
lab435.81.13

Sample 6lab1372.42.770.70.000
lab237661.6
lab3287.716.45.7
lab4282.733.611.9

Sample 7lab189062.470.000
lab2848.722.72.7
lab366954.18.1
lab4611.351.98.5

Sample 8lab11082393.60.000
lab21180201.7
lab3888.332.73.7
lab4937.328.43

Sample 9lab11346119.18.850.006
lab21383.311.50.8
lab3117060.85.2
lab41041.3128.512.3

gNA: not applicable, fewer than three data points due to at least one measure < LOQ.

Table 4

Accuracies calculated for each laboratory based on 3-HPMA fortified samples at 400 ng/ml (samples 2), 1200 ng/ml (samples 3), and 3600 ng/ml (samples 4)

Accuracies (%)
Fortified sample (ng/ml)Lab1Lab2Lab3Lab4

400 ng/ml10811999101
1200 ng/ml1021109792
3600 ng/ml1001079590

aAverage calculated over three independent measures

Individual value plots for 3-HPMA (blue circles) (ng/ml). A. Value plot for the fortified samples in four laboratories. B. Value plot for authentic urine samples in four laboratories. Missing values were below the LOQ. One way ANOVA for 3-HPMA vs laboratories (lab1, 2, 3, and 4) for each set of samples gNA: not applicable, fewer than three data points due to at least one measure < LOQ. Accuracies calculated for each laboratory based on 3-HPMA fortified samples at 400 ng/ml (samples 2), 1200 ng/ml (samples 3), and 3600 ng/ml (samples 4) aAverage calculated over three independent measures A comparison of the repeatability ('r'), reproducibility ('R'), and coefficient of variation for 3-HPMA demonstrated that within-laboratory variation was consistently lower than between-laboratory variation. The average intra-laboratory CoV was 5%, while the average inter-laboratory CoV was 12.2% (Table 5). The average inter-laboratory coefficient of variation was 7% for the fortified urine samples and 16.2% for the authentic urine samples. These results show close comparability with those observed by Biber and colleagues, where samples spiked with cotinine had an inter-laboratory CoV ranging from 3 to 19%, while a CoV range of 4 to 59% was reported for authentic urine samples of smokers [10].
Table 5

Repeatability, reproducibility, and intra, inter-laboratory coefficient of variation for 3-HPMA between the four participating laboratories

SamplesMean 3-HPMA (ng/ml)arbRcCoV within (%)dCoV between (%)e
1345.813.96.1NAf
242850100.84.28.4
31201142.7261.84.27.8
43528268.4460.32.74.7
5457.624.25.9NAf
633054.9138615
7755143.3363.46.817.2
8102189.1346.43.112.1
91235223.1710.16.520.5

aMean of individual 3-HPMA values for all participating laboratories and for the corresponding sample set

bReproducibility

cRepeatability

dcomposite intralaboratory coefficient of variation

einter-laboratory coefficient of variation for the four participating laboratories

fNA: not applicable due to data < LOQ

Repeatability, reproducibility, and intra, inter-laboratory coefficient of variation for 3-HPMA between the four participating laboratories aMean of individual 3-HPMA values for all participating laboratories and for the corresponding sample set bReproducibility cRepeatability dcomposite intralaboratory coefficient of variation einter-laboratory coefficient of variation for the four participating laboratories fNA: not applicable due to data < LOQ The overall average inter-laboratory coefficient of variation for all samples in this study was 12.2%. A CoV value higher than 10% might indicate that there is still some room for improvement; however, this seems to be in line with WHO standardized clinical methods, which in previous studies have reported average inter-laboratory coefficients of variation (CoV) above 10% [11-13]. The results from this first inter-laboratory comparison for the measurement of 3-HPMA in urine demonstrate a reasonably good consensus between laboratories, with an average CoV of 12.2%. However, some consistent measurement biases were still observed between laboratories, suggesting that additional work may be required to reduce the inter-laboratory CoV even further.

Abbreviations

3-HPMA: 3-hydroxypropylmercapturic acid; CoV: coefficient of variation; ESI: electrospray; LC-MS: liquid chromatography-mass spectrometry; r: repeatability; R: reproducibility; TobReg: tobacco product regulation; UV: ultraviolet; WHO: world health organization

Competing interests

British American Tobacco (BAT) funded and designed the study. The participating laboratories are service providers for Biomarkers analysis and have previously conducted biomarkers analyses for BAT.

Authors' contributions

EM, MM, FC, and GS designed the study. GS, KN, BB, and MS conducted the sample analyses. GE performed the statistical analysis. EM and FC drafted the manuscript. All authors read and approved the final manuscript. Written informed consent was obtained from the urine donors for publication of this manuscript and accompanying images. A copy of the form was made available for review by the by the Editor-in-Chief of this journal.

Additional file 1

Authentic urine 3-HPMA concentrations recalculated using a calibration curve derived from the urine samples spiked with synthetic 3-HPMA. The 3-HPMA fortified non-smokers urine samples were used to establish 3-HPMA calibration curves for each laboratory. The concentration of 3-HPMA in authentic smokers urine was back-calculated based on the calibration curve and peak areas. The values were compared with the values obtained by each lab with their own calibration curve. Click here for file
  9 in total

1.  Relationship between machine-derived smoke yields and biomarkers in cigarette smokers in Germany.

Authors:  Gerhard Scherer; Johannes Engl; Michael Urban; Gerhard Gilch; Dinamis Janket; Kirsten Riedel
Journal:  Regul Toxicol Pharmacol       Date:  2006-10-10       Impact factor: 3.271

2.  Requirements for reference (calibration) laboratories in laboratory medicine.

Authors:  Lothar Siekmann
Journal:  Clin Biochem Rev       Date:  2007-11

3.  Determination of nicotine and cotinine in human serum and urine: an interlaboratory study.

Authors:  A Biber; G Scherer; I Hoepfner; F Adlkofer; W D Heller; J E Haddow; G J Knight
Journal:  Toxicol Lett       Date:  1987-01       Impact factor: 4.372

4.  International Standard for serum vitamin B(12) and serum folate: international collaborative study to evaluate a batch of lyophilised serum for B(12) and folate content.

Authors:  Susan J Thorpe; Alan Heath; Sheena Blackmore; Anne Lee; Malcolm Hamilton; Sean O'broin; Bryant C Nelson; Christine Pfeiffer
Journal:  Clin Chem Lab Med       Date:  2007       Impact factor: 3.694

Review 5.  Acrolein: sources, metabolism, and biomolecular interactions relevant to human health and disease.

Authors:  Jan F Stevens; Claudia S Maier
Journal:  Mol Nutr Food Res       Date:  2008-01       Impact factor: 5.914

6.  Interlaboratory comparability of serum cotinine measurements at smoker and nonsmoker concentration levels: a round-robin study.

Authors:  John T Bernert; Peyton Jacob; David B Holiday; Neal L Benowitz; Connie S Sosnoff; Mira V Doig; Colin Feyerabend; Kenneth M Aldous; Mehran Sharifi; Mark D Kellogg; Loralie J Langman
Journal:  Nicotine Tob Res       Date:  2009-11-23       Impact factor: 4.244

7.  Accuracy and reproducibility of automated estradiol-17beta and progesterone assays using native serum samples: results obtained in the Belgian external assessment scheme.

Authors:  Wim Coucke; Nicole Devleeschouwer; Jean-Claude Libeer; Johan Schiettecatte; Manou Martin; Johan Smitz
Journal:  Hum Reprod       Date:  2007-12       Impact factor: 6.918

8.  Quantitation of acrolein-derived (3-hydroxypropyl)mercapturic acid in human urine by liquid chromatography-atmospheric pressure chemical ionization tandem mass spectrometry: effects of cigarette smoking.

Authors:  Steven G Carmella; Menglan Chen; Yan Zhang; Siyi Zhang; Dorothy K Hatsukami; Stephen S Hecht
Journal:  Chem Res Toxicol       Date:  2007-06-09       Impact factor: 3.739

9.  Mandated lowering of toxicants in cigarette smoke: a description of the World Health Organization TobReg proposal.

Authors:  D M Burns; E Dybing; N Gray; S Hecht; C Anderson; T Sanner; R O'Connor; M Djordjevic; C Dresler; P Hainaut; M Jarvis; A Opperhuizen; K Straif
Journal:  Tob Control       Date:  2008-04       Impact factor: 7.552

  9 in total
  2 in total

1.  Urinary biomarkers of smokers' exposure to tobacco smoke constituents in tobacco products assessment: a fit for purpose approach.

Authors:  Evan O Gregg; Emmanuel Minet; Michael McEwan
Journal:  Biomarkers       Date:  2013-07-31       Impact factor: 2.658

2.  Longitudinal stability in cigarette smokers of urinary biomarkers of exposure to the toxicants acrylonitrile and acrolein.

Authors:  Menglan Chen; Steven G Carmella; Chistopher Sipe; Joni Jensen; Xianghua Luo; Chap T Le; Sharon E Murphy; Neal L Benowitz; F Joseph McClernon; Ryan Vandrey; Sharon S Allen; Rachel Denlinger-Apte; Paul M Cinciripini; Andrew A Strasser; Mustafa al'Absi; Jason D Robinson; Eric C Donny; Dorothy Hatsukami; Stephen S Hecht
Journal:  PLoS One       Date:  2019-01-04       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.