Improved motion-compensated image reconstruction for PET using sensitivity correction per respiratory gate and an approximate tube-of-response backprojector.

Nikolaos Dikaios1, Tim D Fryer.   

Abstract

PURPOSE: One limitation of positron emission tomography (PET) imaging of the torso is patient motion. Motion-compensated image reconstruction (MCIR) is one method employed to reduce the deleterious effects of motion. Existing MCIR algorithms use a single sensitivity correction term, which provides inexact normalization for multigate data. Consequently, in this study, the authors derive and examine the performance of an MCIR algorithm with sensitivity correction per gate. In addition, they demonstrate an approximate tube-of-response (TOR) backprojector.
METHODS: Simulated data from the NCAT phantom with six lesions added were used to compare MCIR algorithms with and without the incorporation of sensitivity correction per gate and TOR backprojection to postreconstruction registration (PRR) and images reconstructed without motion correction. To make the simulations more realistic, intragate motion was included. Deformation fields were determined from NCAT anatomical images using a free-form deformation approach with bending energy regularization.
RESULTS: Sensitivity correction per gate and TOR backprojection improved mean lesion contrast-to-noise ratio by 6%-8%, with the maximum increase (21%-23%) found for the smallest lesion. These increases were obtained despite a small increase (3%) in noise as measured by standard deviation in a uniform lung region. Sensitivity correction per gate comes at no extra computational cost, whilst replacing line-of-response backprojection with TOR backprojection increased the overall computation time by ∼20%. In addition, MCIR was found to be superior to PRR, with one factor contributing to this difference being the differential impact of interpolation following deformation. MCIR was also shown to exhibit super-resolution.
CONCLUSIONS: Replacing a single sensitivity correction term in MCIR with sensitivity correction per gate improves lesion detectability. For a small increase in computational expense, further improvements are achieved using an approximate TOR backprojector rather than line-of-response backprojection.

Entities:  

Mesh:

Year:  2011        PMID: 21978040     DOI: 10.1118/1.3611041

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  4D numerical observer for lesion detection in respiratory-gated PET.

Authors:  Auranuch Lorsakul; Quanzheng Li; Cathryn M Trott; Christopher Hoog; Yoann Petibon; Jinsong Ouyang; Andrew F Laine; Georges El Fakhri
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

2.  Development of a 4D Digital Phantom for Tracer Kinetic Modeling and Analysis of Dynamic Perfusion PET and SPECT Simulation Studies.

Authors:  George S K Fung; Takahiro Higuchi; Min Jae Park; W Paul Segars; Benjamin M W Tsui
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2011-10

3.  Non-Rigid Event-by-Event Continuous Respiratory Motion Compensated List-Mode Reconstruction for PET.

Authors:  Chung Chan; John Onofrey; Yiqiang Jian; Mary Germino; Xenophon Papademetris; Richard E Carson; Chi Liu
Journal:  IEEE Trans Med Imaging       Date:  2017-10-10       Impact factor: 10.048

Review 4.  Synergistic motion compensation strategies for positron emission tomography when acquired simultaneously with magnetic resonance imaging.

Authors:  Irene Polycarpou; Georgios Soultanidis; Charalampos Tsoumpas
Journal:  Philos Trans A Math Phys Eng Sci       Date:  2021-07-05       Impact factor: 4.226

  4 in total