Literature DB >> 21975850

A new approach for the pistoning measurement in transtibial prosthesis.

Hossein Gholizadeh1, Noor Azuan Abu Osman, Ása Gulaug Lúvíksdóttir, Arezoo Eshraghi, Mojtaba Kamyab, Wan Abu Bakar Wan Abas.   

Abstract

BACKGROUND: Good suspension lessens the pistoning (vertical displacement) of the residual limb inside the prosthetic socket. Several methods are used for measuring the pistoning. STUDY
DESIGN: Clinical trial.
OBJECTIVES: To introduce and evaluate a new simple method for measuring the pistoning between the soft liner and socket in transtibial prostheses.
METHODS: Five transtibial prostheses with Iceross silicone liner and shuttle lock were made for the subjects. The pistoning was measured between the liner and socket by a photographic method in single limb support on the prosthetic limb (full weight bearing), non-weight bearing and under three static axial loading conditions (30, 60 and 90 N).
RESULTS: This new method enabled us to measure the pistoning between the liner and prosthetic socket. The reproducibility of measurements in different trials of one session and between two sessions by two observers was shown to be high. The average of displacement between the liner and socket was about 9 mm in non-weight bearing compared with full weight bearing. Furthermore, as we expected, the average of pistoning increased consistently by adding the loads.
CONCLUSIONS: This new method provides for easy and inexpensive determination of pistoning between the liner and socket by every prosthetist in clinical settings.

Entities:  

Mesh:

Year:  2011        PMID: 21975850     DOI: 10.1177/0309364611423130

Source DB:  PubMed          Journal:  Prosthet Orthot Int        ISSN: 0309-3646            Impact factor:   1.895


  4 in total

1.  A simulation-based analysis of the effects of variable prosthesis stiffness on interface dynamics between the prosthetic socket and residual limb.

Authors:  Michael A McGeehan; Peter G Adamczyk; Kieran M Nichols; Michael E Hahn
Journal:  J Rehabil Assist Technol Eng       Date:  2022-07-15

2.  Systematic Experimental Assessment of a 2D-Motion Sensor to Detect Relative Movement between Residual Limb and Prosthetic Socket.

Authors:  Veronika Noll; Stephan Rinderknecht; Philipp Beckerle
Journal:  Sensors (Basel)       Date:  2018-07-06       Impact factor: 3.576

3.  Development of an air pneumatic suspension system for transtibial prostheses.

Authors:  Gholamhossein Pirouzi; Noor Azuan Abu Osman; Azim Ataollahi Oshkour; Sadeeq Ali; Hossein Gholizadeh; Wan A B Wan Abas
Journal:  Sensors (Basel)       Date:  2014-09-09       Impact factor: 3.576

4.  Silicone soft socket system for the treatment of geriatric transtibial amputees.

Authors:  Robert Breuer; Klemens Trieb
Journal:  Wien Klin Wochenschr       Date:  2020-04-16       Impact factor: 1.704