Effect of mini-implant diameter on fracture risk and self-drilling efficacy.

Sérgio Estelita Barros1, Guilherme Janson, Kelly Chiqueto, Daniela Gamba Garib, Marcos Janson.   

Abstract

INTRODUCTION: The aim of this study was to evaluate the effect of mini-implant diameter on fracture risk and self-drilling efficacy.
METHODS: A sample of 405 mini-implants with 9 diameters from 1.2 to 2.0 mm was used. Ten mini-implants of each diameter were placed in artificial bone, and 25 were placed in pig iliac bone to evaluate placement torque (PT) and axial placement load (APL), which represents self-drilling efficacy. Ten mini-implants of each diameter were used to determine fracture torque (FT). The different diameters were compared regarding PT, FT, and APL. The fracture risk of each diameter was evaluated by the fracture resistance index (FT/PT × [FT-PT]). The PT and APL changes during placement were correlated.
RESULTS: Only PT and FT were different for all mini-implant diameter changes. PT and FT showed a strong correlation with the mini-implant diameter, but the APL was weakly to moderately correlated. The fracture resistance index was remarkably greater for each 0.1 mm added in diameter. The PT increased significantly, whereas the APL was progressively reduced during placement.
CONCLUSIONS: Increases in mini-implant diameters significantly influenced the increases of PT and FT on quantities that progressively reduced the fracture risk. The self-drilling efficacy was not strongly influenced by diameter.
Copyright © 2011. Published by Mosby, Inc.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21967956     DOI: 10.1016/j.ajodo.2011.06.016

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  13 in total

1.  Surgical management of fractured orthodontic mini- implant- a case report.

Authors:  Manthan Desai; Anoop Jain; Nida Sumra
Journal:  J Clin Diagn Res       Date:  2015-01-01

2.  Impact of mini-implant length on stability at the initial healing period: a controlled clinical study.

Authors:  Manuel Nienkemper; Benedict Wilmes; Alexander Pauls; Dieter Drescher
Journal:  Head Face Med       Date:  2013-10-20       Impact factor: 2.151

3.  Stress distribution patterns at mini-implant site during retraction and intrusion--a three-dimensional finite element study.

Authors:  Gautham Sivamurthy; Shantha Sundari
Journal:  Prog Orthod       Date:  2016-01-18       Impact factor: 2.750

4.  Fracture strength of orthodontic mini-implants.

Authors:  Tatiana Feres Assad-Loss; Flávia Mitiko Fernandes Kitahara-Céia; Giordani Santos Silveira; Carlos Nelson Elias; José Nelson Mucha
Journal:  Dental Press J Orthod       Date:  2017 May-Jun

5.  Influence of Screw Length and Bone Thickness on the Stability of Temporary Implants.

Authors:  Daniel Jogaib Fernandes; Carlos Nelson Elias; Antônio Carlos de Oliveira Ruellas
Journal:  Materials (Basel)       Date:  2015-09-23       Impact factor: 3.623

6.  An in vitro study of factors affecting the primary stability of orthodontic mini-implants.

Authors:  Lindsy Holm; Susan J Cunningham; Aviva Petrie; Richard R J Cousley
Journal:  Angle Orthod       Date:  2012-05-07       Impact factor: 2.079

7.  Effects of insertion angle and implant thread type on the fracture properties of orthodontic mini-implants during insertion.

Authors:  Il-Sik Cho; Tae-Woo Kim; Sug-Joon Ahn; Il-Hyung Yang; Seung-Hak Baek
Journal:  Angle Orthod       Date:  2012-12-10       Impact factor: 2.079

8.  Design characteristics, primary stability and risk of fracture of orthodontic mini-implants: pilot scan electron microscope and mechanical studies.

Authors:  André Walter; Heinz Winsauer; Jordi Marcé-Nogué; Sergi Mojal; Andreu Puigdollers
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2013-09-01

9.  Torque Loss After Miniscrew Placement: An In-Vitro Study Followed by a Clinical Trial.

Authors:  Marco Migliorati; Sara Drago; Fabrizio Barberis; Irene Schiavetti; Domenico Dalessandri; Stefano Benedicenti; Armando Silvestrini Biavati
Journal:  Open Dent J       Date:  2016-05-31

10.  Effectiveness of methods for detaching orthodontic implants likely to fracture upon rotational torque - an animal study.

Authors:  Susumu Nakagaki; Masahiro Iijima; Yoshitaka Yasuda; Keisuke Handa; Toshiyuki Koike; Takashi Saito; Itaru Mizoguchi
Journal:  Clin Exp Dent Res       Date:  2016-02-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.