Determination of RW3-to-water mass-energy absorption coefficient ratio for absolute dosimetry.

Katrina Y T Seet1, Peta M Hanlon, Paul H Charles.   

Abstract

The measurement of absorbed dose to water in a solid-phantom may require a conversion factor because it may not be radiologically equivalent to water. One phantom developed for the use of dosimetry is a solid water, RW3 white-polystyrene material by IBA. This has a lower mass-energy absorption coefficient than water due to high bremsstrahlung yield, which affects the accuracy of absolute dosimetry measurements. In this paper, we demonstrate the calculation of mass-energy absorption coefficient ratios, relative to water, from measurements in plastic water and RW3 with an Elekta Synergy linear accelerator (6 and 10 MV photon beams) as well as Monte Carlo modeling in BEAMnrc and DOSXYZnrc. From this, the solid-phantom-to-water correction factor was determined for plastic water and RW3.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21960410     DOI: 10.1007/s13246-011-0102-4

Source DB:  PubMed          Journal:  Australas Phys Eng Sci Med        ISSN: 0158-9938            Impact factor:   1.430


  2 in total

1.  Dosimetry audits in Taiwan radiotherapy departments.

Authors:  An-Cheng Shiau; Shih-Ming Hsu; Pei-Yun Huang; Chiu-Ping Chen; Yi-Ting Huang; Ke-Yu Lien; Chin-Ping Chen; Shung-Hwei Fan; Shiu-Chen Jeng; Ho-Hsing Chen; Ji-An Liang
Journal:  BJR Open       Date:  2021-04-15

2.  Dose comparison between Gafchromic film, XiO, and Monaco treatment planning systems in a novel pelvic phantom that contains a titanium hip prosthesis.

Authors:  Nicholas Ade; F C P du Plessis
Journal:  J Appl Clin Med Phys       Date:  2017-07-25       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.