Radiation dose in coronary artery angiography with 320-detector row CT and its diagnostic accuracy: comparison with 64-detector row CT.

T Zhang1, Z Luo, D Wang, D Han, J Bai, X Meng, B Shen.   

Abstract

AIM: The aim of this study was the comparison of image quality, diagnostic accuracy and radiation dose of ECG-gated volume scan mode of 320-detector row CT with 64-detector row CT.
METHODS: A hundred and twenty patients receiving coronary artery CT angiography (CTA) were divided into two groups. Sixty patients in Group A underwent 64-detector row CTA; for the remainder, Group B, 320-detector row CT was used. Conventional coronary angiography was then performed in all patients. In Group B, patients with a pulse <65 beat/min underwent a prospective ECG-gated volume scan using a 70-80% R-R phase window and 1-beat, whereas a 30-80% R-R phase window, 2-beat CT was performed in patients with a pulse >65 beat/min.
RESULTS: In Group A, 247 vessels and 920 segments were assessed; in Group B, 242 and 906. In 320-detector row CTA, the coronary stenosis diagnosis' specificities were 98%, 96%, and 95% at the segmental, vessel, and patient levels, respectively (98%, 94%, and 92% in 64-detector row CTA); the negative predictive values were 99%, 98%, and 100% (98%, 97%, and 95% in 64-detector row CTA). There was no significant difference in specificity or negative predictive values between the groups. The average effective radiation dose in Group B was lower than in Group A (P<0.001). In Group B, the dose for 48 patients with 70-80% R-R phase window, 1-beat scan, was lower than for the remaining 12 patients (30-80% R-R phase window, 2-beat) (P<0.001).
CONCLUSION: The 320-detector CT reduces the radiation exposure by approximately 80% compared with 64-detector CT, with similar diagnostic accuracy and image quality.

Entities:  

Mesh:

Year:  2011        PMID: 21959699

Source DB:  PubMed          Journal:  Minerva Med        ISSN: 0026-4806            Impact factor:   4.806


  4 in total

1.  Improved evaluation of calcified segments on coronary CT angiography: a feasibility study of coronary calcium subtraction.

Authors:  Ryoichi Tanaka; Kunihiro Yoshioka; Kenta Muranaka; Takuya Chiba; Takanori Ueda; Tadashi Sasaki; Tetsuya Fusazaki; Shigeru Ehara
Journal:  Int J Cardiovasc Imaging       Date:  2013-10-25       Impact factor: 2.357

Review 2.  Meta-analysis: diagnostic accuracy of coronary CT angiography with prospective ECG gating based on step-and-shoot, Flash and volume modes for detection of coronary artery disease.

Authors:  Linfeng Yang; Tao Zhou; Ruijie Zhang; Lin Xu; Zhaohui Peng; Juan Ding; Sen Wang; Min Li; Gang Sun
Journal:  Eur Radiol       Date:  2014-05-28       Impact factor: 5.315

3.  Preliminary study of prospective ECG-gated 320-detector CT coronary angiography in patients with ventricular premature beats.

Authors:  Tong Zhang; Jinquan Bai; Wei Wang; Dan Wang; Baozhong Shen
Journal:  PLoS One       Date:  2012-06-06       Impact factor: 3.240

Review 4.  The diagnostic performance of coronary artery angiography with 64-MSCT and post 64-MSCT: systematic review and meta-analysis.

Authors:  Min Li; Xiang-Min Du; Zhi-Tao Jin; Zhao-Hui Peng; Juan Ding; Li Li
Journal:  PLoS One       Date:  2014-01-21       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.