Volumetric MR-HIFU ablation of uterine fibroids: role of treatment cell size in the improvement of energy efficiency.

Young-sun Kim1, Bilgin Keserci, Ari Partanen, Hyunchul Rhim, Hyo K Lim, Min Jung Park, Max O Köhler.   

Abstract

PURPOSE: To evaluate the energy efficiency of differently sized volumetric ablations in MR-guided high-intensity focused ultrasound (MR-HIFU) treatment of uterine fibroids.
MATERIALS AND METHODS: This study was approved by the institutional review board and informed consent was obtained from all participants. Ten symptomatic uterine fibroids (mean diameter 8.9 cm) in 10 women (mean age 42.2) were treated by volumetric MR-HIFU ablation under binary feedback control. The energy efficiency (mm3/J) of each sonication was calculated as the volume of lethal thermal dose (240 equivalent minutes at 43 °C) per unit acoustic energy applied. Operator-controllable parameters and signal intensity ratio of uterine fibroid to skeletal muscle on T2-weighted MR images were tested with univariate and multivariate analyses to discern which parameters significantly correlated with the ablation energy efficiency.
RESULTS: We analyzed a total of 236 sonications. The energy efficiency of the ablations was 0.42±0.25 mm3/J (range 0.004-1.18) with energy efficiency improving with the treatment cell size (4 mm, 0.06±0.06 mm3/J; 8 mm, 0.29±0.12 mm3/J; 12 mm, 0.58±0.18 mm3/J; 16 mm, 0.91±0.17 mm3/J). Treatment cell size (r=0.814, p<0.001), distance of ultrasound propagation (r=-0.151, p=0.020), sonication frequency (1.2 or 1.45 MHz; p<0.001), and signal intensity ratio (r=-0.205, p=0.002) proved to be significant by univariate analysis, while multivariate analysis revealed treatment cell size (B=0.075, p<0.001), US propagation distance (B=-6.928, p<0.001), and signal intensity ratio (B=-0.024, p=0.001) to be independently significant.
CONCLUSION: Energy efficiency in volumetric MR-HIFU ablation of uterine fibroids improves with increased treatment cell size, independent of other significant contributors such as distance of ultrasound propagation or signal intensity of the tumor on T2-weighted MR imaging.
Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21959213     DOI: 10.1016/j.ejrad.2011.09.005

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  29 in total

Review 1.  Reproductive impact of MRI-guided focused ultrasound surgery for fibroids: a systematic review of the evidence.

Authors:  Natalie A Clark; Sunni L Mumford; James H Segars
Journal:  Curr Opin Obstet Gynecol       Date:  2014-06       Impact factor: 1.927

Review 2.  Review of influential clinical factors in reducing the risk of unsuccessful MRI-guided HIFU treatment outcome of uterine fibroids.

Authors:  Nguyen Minh Duc; Bilgin Keserci
Journal:  Diagn Interv Radiol       Date:  2018-09       Impact factor: 2.630

3.  Supersonic transient magnetic resonance elastography for quantitative assessment of tissue elasticity.

Authors:  Yu Liu; Jingfei Liu; Brett Z Fite; Josquin Foiret; Asaf Ilovitsh; J Kent Leach; Erik Dumont; Charles F Caskey; Katherine W Ferrara
Journal:  Phys Med Biol       Date:  2017-04-20       Impact factor: 3.609

4.  Mild hyperthermia with magnetic resonance-guided high-intensity focused ultrasound for applications in drug delivery.

Authors:  Ari Partanen; Pavel S Yarmolenko; Antti Viitala; Sunil Appanaboyina; Dieter Haemmerich; Ashish Ranjan; Genevieve Jacobs; David Woods; Julia Enholm; Bradford J Wood; Matthew R Dreher
Journal:  Int J Hyperthermia       Date:  2012       Impact factor: 3.914

5.  Magnetic resonance imaging-guided volumetric ablation of symptomatic leiomyomata: correlation of imaging with histology.

Authors:  Aradhana M Venkatesan; Ari Partanen; Tajana Klepac Pulanic; Matthew R Dreher; John Fischer; Robert K Zurawin; Raja Muthupillai; Sham Sokka; Heikki J Nieminen; Ninet Sinaii; Maria Merino; Bradford J Wood; Pamela Stratton
Journal:  J Vasc Interv Radiol       Date:  2012-06       Impact factor: 3.464

6.  Mechanical fractionation of tissues using microsecond-long HIFU pulses on a clinical MR-HIFU system.

Authors:  Avinash Eranki; Navid Farr; Ari Partanen; Karun V Sharma; Christopher T Rossi; Avi Z Rosenberg; AeRang Kim; Matthew Oetgen; Haydar Celik; David Woods; Pavel S Yarmolenko; Peter C W Kim; Bradford J Wood
Journal:  Int J Hyperthermia       Date:  2018-02-22       Impact factor: 3.914

Review 7.  Clinical and Technical Aspects of MR-Guided High Intensity Focused Ultrasound for Treatment of Symptomatic Uterine Fibroids.

Authors:  Laura E Rueff; Steven S Raman
Journal:  Semin Intervent Radiol       Date:  2013-12       Impact factor: 1.513

Review 8.  Magnetic Resonance-Guided Drug Delivery.

Authors:  Andrew S Mikhail; Ari Partanen; Pavel Yarmolenko; Aradhana M Venkatesan; Bradford J Wood
Journal:  Magn Reson Imaging Clin N Am       Date:  2015-07-09       Impact factor: 2.266

9.  Stereotactic Transcranial Focused Ultrasound Targeting System for Murine Brain Models.

Authors:  Sang Won Choi; Tyler I Gerhardson; Sarah E Duclos; Rachel K Surowiec; Ulrich M Scheven; Stefanie Galban; Fred T Lee; Joan M Greve; James M Balter; Timothy L Hall; Zhen Xu
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2020-12-23       Impact factor: 2.725

10.  Analytical estimation of ultrasound properties, thermal diffusivity, and perfusion using magnetic resonance-guided focused ultrasound temperature data.

Authors:  C R Dillon; G Borasi; A Payne
Journal:  Phys Med Biol       Date:  2016-01-07       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.