Validity and ease of use of a computerized Hess chart.

Patrick Watts1, Harish Nayak, May Kim Lim, Alan Ashcroft, Hasan Al Madfai, Hollie Palmer.   

Abstract

PURPOSE: To investigate agreement between computerized and conventional methods for obtaining Hess charts and to compare relative ease of use of both methods.
METHODS: Hess charts of 65 patients were obtained by the use of the computerized Assaf Ocular Motility Analyzer (OMA) and the conventional Lees screen method. The Hess charts produced by each method were compared with a previously described scoring system. Patients compared the ease of testing by using a 5-point Likert scale.
RESULTS: For horizontal deviations of the right eye, the OMA provided a significantly larger (P = 0.0001) deviation (301° ± 267°) than the Lees screen (204° ± 306°). The Lees screen gave a significantly larger score for vertical deviations of the left eye (117° ± 158° vs 96° ± 129°; P = 0.003). Vertical deviations of the right eye and horizontal deviations of the left eye did not differ significantly between tests. Patients required the same amount of time to complete both tests, but the OMA was slightly easier to perform than the Lees screen (Likert score, 1.2 ± 0.5 vs 1.3 ± 0.4; P = 0.046).
CONCLUSIONS: The scores measured were larger with OMA in the horizontal and smaller in the vertical direction than with the Lees screen in some directions. Although the OMA did not save time, patients found it easier to perform than the Lees screen. The OMA may be considered a useful alternative to the Lees screen.
Copyright © 2011 American Association for Pediatric Ophthalmology and Strabismus. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21958903     DOI: 10.1016/j.jaapos.2010.12.021

Source DB:  PubMed          Journal:  J AAPOS        ISSN: 1091-8531            Impact factor:   1.220


  3 in total

1.  Adult reference values of the computerized diplopia test.

Authors:  Ling-Yun Zhou; Tie-Juan Liu; Xue-Mei Li
Journal:  Int J Ophthalmol       Date:  2016-11-18       Impact factor: 1.779

2.  Stepping into the virtual unknown: feasibility study of a virtual reality-based test of ocular misalignment.

Authors:  N Nesaratnam; P Thomas; A Vivian
Journal:  Eye (Lond)       Date:  2017-06-02       Impact factor: 3.775

3.  Quantifying Eye Alignment in Orbital Fracture Patients: The Digital Hess Screen.

Authors:  Sarah R Akkina; Aliya Shabbir; Annamarie Lahti; Raghu C Mudumbai; Christopher B Chambers; Kris S Moe; James O Phillips
Journal:  Facial Plast Surg Aesthet Med       Date:  2020-05-22
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.