Ventricular activation sequence during left ventricular pacing promotes QRS complex oversensing in the atrial channel.

Marek Jastrzebski1.   

Abstract

BACKGROUND: Left ventricular (LV)-only pacing has a significant effect on delay in depolarization of parts of the ventricles that are likely oversensed in the right atrial channel. The study aimed to assess the impact of ventricular activation sequence on QRS oversensing and far-field endless-loop pacemaker tachycardia (ELT) in patients who received cardiac resynchronization therapy (CRT) devices.
METHODS: The study examined 102 patients with CRT devices. Oversensing artifacts in the atrial channel were inspected on intracardiac electrograms, and their timing with respect to the beginning of QRS was determined during DDD-right ventricular (RV), DDD-LV, DDD-biventricular (BiV), and AAI pacing modes. The occurrence of ELT during DDD-LV pacing with a postventricular atrial refractory period (PVARP) of 250 ms was also assessed.
RESULTS: The timing of oversensing artifacts (in relation to the beginning of surface QRS) was dependent on ventricular activation sequence, occurring promptly following intrinsic activation via the right bundle branch (47.1 ± 26.4 ms), later during RV pacing (108.7 ± 22.5 ms) or BiV pacing (109.4 ± 23.1 ms), and significantly later, corresponding to the final part of the QRS, during LV pacing (209.6 ± 40.0 ms, range: 140-340 ms, P < 0.001). Oversensing was significantly more frequent during LV than during RV pacing (35.3% vs 22.5%, P < 0.001). Far-field ELT was observed in six patients.
CONCLUSIONS: Oversensing artifacts in the atrial channel are likely caused by depolarization of the basal part of the right ventricle. The novel mechanism of QRS oversensing outside PVARP, caused by a reversed ventricular activation sequence during LV-only pacing, may be important in some CRT patients. ©2011, The Author. Journal compilation ©2011 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21954967     DOI: 10.1111/j.1540-8159.2011.03219.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Use of an atrial lead with very short tip-to-ring spacing avoids oversensing of far-field R-wave.

Authors:  Christof Kolb; Georg Nölker; Carsten Lennerz; Hansmartin Jetter; Verena Semmler; Klaus Pürner; Klaus-Jürgen Gutleben; Tilko Reents; Klaus Lang; Ulrich Lotze
Journal:  PLoS One       Date:  2012-06-22       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.