Literature DB >> 21950720

Clinical evaluation of a simultaneous closed-loop anaesthesia control system for depth of anaesthesia and neuromuscular blockade*.

M Janda1, O Simanski, J Bajorat, B Pohl, G F E Noeldge-Schomburg, R Hofmockel.   

Abstract

We developed a closed-loop system to control the depth of anaesthesia and neuromuscular blockade using the bispectral index and the electromyogram simultaneously and evaluated the clinical performance of this combined system for general anaesthesia. Twenty-two adult patients were included in this study. Anaesthesia was induced by a continuous infusion of remifentanil at 0.4 μg.kg(-1) .min(-1) (induction dose) and then 0.25 μg.kg(-1) .min(-1) (maintenance dose) and propofol at 2 mg.kg(-1) 3 min later. The combined automatic control was started 2 min after tracheal intubation. The depth of anaesthesia was recorded using bispectral index monitoring using a target value of 40. The target value of neuromuscular blockade, using mivacurium, was a T1/T1(0) twitch height of 10%. The precision of the system was calculated using internationally defined performance parameters. Twenty patients were included in the data analysis. The mean (SD) duration of simultaneous control was 129 (69) min. No human intervention was necessary during the computer-controlled administration of propofol and mivacurium. All patients assessed the quality of anaesthesia as 'good' to 'very good'; there were no episodes of awareness. The mean (SD) median performance error, median absolute performance error and wobble for the control of depth of anaesthesia and for neuromuscular blockade were -0.31 (1.78), 6.76 (3.45), 6.32 (2.93) and -0.38 (1.68), 3.75 (4.83), 3.63 (4.69), respectively. The simultaneous closed-loop system using propofol and mivacurium was able to maintain the target values with a high level of precision in a clinical setting.
© 2011 The Authors. Anaesthesia © 2011 The Association of Anaesthetists of Great Britain and Ireland.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21950720     DOI: 10.1111/j.1365-2044.2011.06875.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  9 in total

1.  Rocuronium: automatic infusion versus manual administration with TOF monitorisation.

Authors:  Fatma Gulcin Ozturk Arikan; Guldem Turan; Asu Ozgultekin; Zubeyir Sivrikaya; Bekir Cem Cosar; Dondu Nisa Onder
Journal:  J Clin Monit Comput       Date:  2015-08-18       Impact factor: 2.502

2.  Optimizing Robust PID Control of Propofol Anesthesia for Children: Design and Clinical Evaluation.

Authors:  Klaske van Heusden; Kristian Soltesz; Erin Cooke; Sonia Brodie; Nicholas West; Matthias Gorges; J Mark Ansermino; Guy A Dumont
Journal:  IEEE Trans Biomed Eng       Date:  2019-02-08       Impact factor: 4.538

Review 3.  Machine Learning and Artificial Intelligence in Neurocritical Care: a Specialty-Wide Disruptive Transformation or a Strategy for Success.

Authors:  Fawaz Al-Mufti; Michael Kim; Vincent Dodson; Tolga Sursal; Christian Bowers; Chad Cole; Corey Scurlock; Christian Becker; Chirag Gandhi; Stephan A Mayer
Journal:  Curr Neurol Neurosci Rep       Date:  2019-11-13       Impact factor: 5.081

4.  Evaluation of SLOG/TCI-III pediatric system on target control infusion of propofol.

Authors:  Wan-hua Yang; Hong-bin Gu; Bing Chen; Juan Li; Qiu-wei Fan; Yong-fang Yuan; Xiangdong Wang
Journal:  J Transl Med       Date:  2011-11-01       Impact factor: 5.531

Review 5.  A Review of Modern Control Strategies for Clinical Evaluation of Propofol Anesthesia Administration Employing Hypnosis Level Regulation.

Authors:  Muhammad Ilyas; Muhammad Fasih Uddin Butt; Muhammad Bilal; Khalid Mahmood; Ali Khaqan; Raja Ali Riaz
Journal:  Biomed Res Int       Date:  2017-03-30       Impact factor: 3.411

6.  Cost-effectiveness of anesthesia maintained with sevoflurane or propofol with and without additional monitoring: a prospective, randomized controlled trial.

Authors:  Timea Bocskai; Csaba Loibl; Zoltan Vamos; Gabor Woth; Tihamer Molnar; Lajos Bogar; Laszlo Lujber
Journal:  BMC Anesthesiol       Date:  2018-07-28       Impact factor: 2.217

7.  Study of the rational dose of propofol in elderly patients under bispectral index monitoring during total intravenous anesthesia: A PRISMA-compliant systematic review.

Authors:  Lina Jia; Jiachen Hou; Haibo Zheng; Lihua Sun; Yingying Fan; Xu Wang; Mingyue Hao; Yue Li; Tongwei Yang
Journal:  Medicine (Baltimore)       Date:  2020-01       Impact factor: 1.889

8.  Fully automated life support: an implementation and feasibility pilot study in healthy pigs.

Authors:  Wilfried Klingert; Jörg Peter; Christian Thiel; Karolin Thiel; Wolfgang Rosenstiel; Kathrin Klingert; Christian Grasshoff; Alfred Königsrainer; Martin Schenk
Journal:  Intensive Care Med Exp       Date:  2018-01-16

9.  Hypnosis regulation in propofol anaesthesia employing super-twisting sliding mode control to compensate variability dynamics.

Authors:  Muhammad Ilyas; Jamshed Iqbal; Sayyar Ahmad; Ali Arshad Uppal; Waqas Ahmad Imtiaz; Raja Ali Riaz
Journal:  IET Syst Biol       Date:  2020-04       Impact factor: 1.615

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.