| Literature DB >> 21947202 |
Lei Zhang1, Pan Wang, Yao Xiao, Huakang Yu, Limin Tong.
Abstract
We report a microfibre absorption sensor by using a 900 nm diameter silica microfibre embedded in a 125 μm wide microchannel with a detection length of 2.5 cm. Investigated by measuring the absorbance of methylene blue (MB), the sensor shows a detection limit down to 50 pM with excellent reversibility in a concentration range of 0-5 nM. The sensor has also been applied to bovine serum albumin (BSA) measurement, with a detection limit of 10 fg mL(-1). In addition, the sample volume requirement is merely 500 nL with a probing light power of about 150 nW, which is very promising for safe detection of single or a few molecules of biological specimens. This journal is © The Royal Society of Chemistry 2011Entities:
Year: 2011 PMID: 21947202 DOI: 10.1039/c1lc20519f
Source DB: PubMed Journal: Lab Chip ISSN: 1473-0189 Impact factor: 6.799