Artiflex toric phakic intraocular lens implantation in congenital nystagmus.

Gonzalo Muñoz1, Lurdes Belda, César Albarrán-Diego, Teresa Ferrer-Blasco, Santiago García-Lázaro.   

Abstract

CASE: A 44-year-old woman with congenital nystagmus and myopic astigmatism in both eyes who was submitted to phakic intraocular lens (pIOL) implantation.
METHODS: Full ophthalmologic examination including refractive status, corrected (CDVA) and uncorrected (UCVA) monocular and binocular visual acuities, ocular motility, slit-lamp evaluation, tonometry and fundoscopy before and after implantation of toric pIOLs (Artiflex; Ophtec BV, The Netherlands) in both eyes.
RESULTS: Preoperative logMAR CDVA were 0.699 and 0.420 in the right and left eye, respectively. Three months after surgery, logMAR UCVA were 0.398 and 0.182, reaching binocular logMAR UCVA of 0.132. There were no changes in nystagmus characteristics after surgery.
CONCLUSIONS: pIOL implantation may be a safe and suitable treatment to correct high refractive errors in patients with congenital nystagmus. Significant improvement in CDVA and UCVA can be achieved.

Entities:  

Keywords:  Nystagmus; Refractive surgery; Toric phakic intraocular lens

Year:  2011        PMID: 21941505      PMCID: PMC3177809          DOI: 10.1159/000331115

Source DB:  PubMed          Journal:  Case Rep Ophthalmol        ISSN: 1663-2699


Introduction

Nystagmus is an ocular motor disorder characterized by involuntary and rhythmic movements of one or both eyes [1]. Nystagmus can be classified according to the age of onset (congenital or acquired), direction (horizontal, vertical, torsional or combinations) or considering the character of movements (jerk, pendular, dual jerk) [1, 2]. Patients with nystagmus usually have a diminished visual performance because of these eye movements. This includes deficiencies in visual acuity inversely proportional to the intensity of nystagmus [3], and anomalies in color vision, contrast sensitivity and stereopsis [2]. The more common clinical signs and symptoms are amblyopia, anomalous head posture, oscillopsia, photophobia and hypoaccommodation [1, 2]. Different ways to correct refractive errors in patients with nystagmus can be found, grouped into non-surgical and surgical procedures. Those related to non-surgical treatments include spectacles [1, 2] and contact lenses [3]. On the other hand, laser corneal refractive surgery has been recently applied as an alternative to treat these patients [4]. However, to the best of our knowledge, no previous reports regarding the use of phakic intraocular lenses (pIOLs) for correcting refractive errors in patients with nystagmus has been reported. The present study reports the case of a patient with congenital nystagmus submitted to Artiflex toric pIOL (Ophtec BV, The Netherlands) implantation for refractive error correction.

Case Report

A 44-year-old woman with congenital nystagmus visited our center aiming to obtain spectacle independence. She had been wearing contact lenses for many years, but she has suffered from dryness for 2 years and discontinued lens wear. Her manifest refraction was –8.50 –3.50 × 20° in her right eye (RE) and –8.00 –3.50 × 160° in her left eye (LE). Corrected distance visual acuity (CDVA) using spectacles was 0.699 logMAR (RE) and 0.420 logMAR (LE). The patient displayed a horizontal jerk nystagmus 10° in amplitude. The nystagmus frequency was moderate and increased with attempted fixation. No well-defined null zone or anomalous head movement or position could be determined. Preoperative examination included manifest refraction, CDVA using spectacles, slit-lamp biomicroscopy, Goldman tonometry, binocular indirect ophthalmoscopy, biometry, pachymetry, keratometry and corneal topography ( ). pIOL implantation surgery was explained, and written informed consent was obtained.
Table 1

Patient characteristics before and after surgery

Right eyeLeft eye
Preoperative
ACD, mm03.55003.55
AL, mm27.24026.67
Keratometry, D
 K142.62042.75
 K246.25046.00
BCVA, logMAR00.699000.420
Refraction, D−8.50 −3.50 ×20°0–8.00 −3.50 ×160°
pIOL power, D−9.00 −3.50 × 0°0–8.50 −3.50 ×0°
Axis pIOL implantation, °20160

Postoperative
UCVA, logMAR00.398000.182
BCVA, logMAR00.357000.155
Refraction, D−0.50 ×60°0–0.50 ×150°

ACD = Anterior chamber depth; AL = axial length; BCVA = best spectacle-corrected visual acuity; pIOL = phakic intraocular lens; UCVA = uncorrected distance visual acuity.

The Artiflex toric pIOL consists of a three-piece lens made up of hydrophobic polysiloxane optic and polymethylmethacrylate rigid haptics. It is an iris-fixated foldable pIOL with a 6-mm optical zone diameter, designed for the correction of regular myopic astigmatism. Refractive power sphere varies from –1.0 to –13.5 diopters (D) in combination with a cylinder correction from –1.0 to –5.0 D. Furthermore, the Artiflex toric pIOL claw design guarantees long-term rotational stability in addition to avoiding endothelium-lens contact. The power of the pIOL was calculated using the van der Heijde formula, which uses vertex distance, subjective spectacle refraction, corneal curvature and anterior chamber depth. Both eyes were treated by the same surgeon (G.M.) within a 24-hour period. Surgery was carried out under peribulbar anesthesia. The Artiflex toric pIOL was implanted through a 3.2-mm clear-cornea incision located at 12 o'clock in both eyes. The marking technique previously reported by Graether [5] was used, consisting of placing a single ink mark at the limbus at 6 o'clock with the patient in the upright sitting position during the preoperative preparation. A 360° gauge was applied onto the eye having the patient lying back, and aligning the 90-270° axis on the mark at the 6 o'clock position previously made to compensate for a postural axis shift. A half-circle axis marker was placed so that the index mark on the semicircle was aligned with the minus axis on the cornea (20-200° in the RE and 160-340° in the LE), which was the axis where the Artiflex toric pIOL had to be fixated. Two 1-mm side port incisions were done at the appropriate axis to fixate the pIOL to the iris by means of forceps. At the end of the procedure, a single 10/0 nylon suture was used to secure the incision in both eyes. There were no complications in the surgical procedures. Postoperative treatment included tobramycin and dexamethasone eyedrops (TobraDex, AlconLabs) 4 times daily for 7 days and then tapering it slowly during the following 3 weeks. Lubricant eyedrops (Acuolens, AlconLabs) were prescribed 3-4 times daily for 3 months. Postoperative examinations included monocular and binocular uncorrected distance visual acuity (UCVA), manifest refraction, CDVA and intraocular pressure (IOP) measurements performed at 1 day, 1 week and 1 and 3 months after surgery. No postoperative complications appeared including high IOP. The slit-lamp examination showed no pIOL rotation, they were well-centered in both eyes ( ). UCVA at 3 months of follow-up were 0.398 (logMAR) and 0.182 (logMAR) for the RE and LE respectively, with binocular UCVA of 0.132 (logMAR), while best spectacle-corrected visual acuity was 0.378 (logMAR) with –0.75 × 60° in her RE and 0.155 (logMAR) with –0.75 × 150° in her LE. Postoperative binocular UCVA at 3 months was 0.132 (logMAR).

Artiflex toric phakic intraocular lens centered in the right and left eye.

Ethics

All study procedures adhered to the recommendations of the declaration of Helsinki.

Discussion

Nystagmus is an ocular motor disorder with repetitive and involuntary oscillations of the eye. It is usually associated to visual dysfunctions and vision impairment including significant amblyopia. Different optical and surgical treatment modalities have been proposed in order to reduce oscillations and improve visual performance [1, 2]. Correction of significant refractive error in patients with nystagmus is the single most powerful intervention to improve visual functions [2]. The use of contact lenses has also been described in patients with nystagmus [3]. Contact lenses correct refractive errors moving simultaneously along with the eye and providing a more continuous fixation than spectacles, reducing higher-order optical aberrations together with the prismatic effect. Allen and Davies [3] evaluated the use of contact lenses in patients with nystagmus, reporting significant improvement of 2 lines of visual acuity and better contrast sensitivity in comparison with spectacles. Recently, laser corneal refractive surgery has been used in patients with nystagmus [4]. As previously stated, visual acuity in nystagmus is often poor due to the extremely short foveation time, making these patients theoretically poor candidates for laser corneal refractive surgery. Previous authors consider nystagmus a contraindication to laser corneal refractive surgery because of the involuntary ocular movements which might cause decentered ablation and problems associated to the suction system for creating the flap. However, new laser systems use high-speed active eye-tracking which may counteract eye motion during laser ablation. In addition, suction systems to fixate the globe during flap preparation seem to be effective to avoid flap-creation problems in these patients. For instance, Mahler et al [4] found good outcomes in 8 patients with nystagmus (CDVA improved in 62.5% of the eyes). They concluded that LASIK surgery may be safely performed in nystagmus using an active tracking system at 120 Hz with or without mechanical fixation. Patients with nystagmus may display high refractive errors that could contraindicate laser corneal surgery, making pIOLs the alternative of treatment. Similarly, a recent report from the American Academy of Ophthalmology concluded that in cases of myopia levels equal to or higher than –8 D, pIOLs may provide a better visual outcome than keratorefractive surgery [6]. Up until now, no reports have been available in the literature considering pIOLs in patients with nystagmus. A case of a young patient with nystagmus aiming to obtain spectacle independency is presented here. This patient was submitted to Artiflex toric pIOL implantation in both eyes. After the surgery, the patient had no visual complaints and achieved an UCVA better than preoperative spectacles CDVA, improving 2 and 3 lines of visual acuity in the RE and LE, respectively. When comparing preoperative and postoperative visual acuity, it should be considered that normally a change in magnification of the retinal image occurs. For myopes who are corrected with spectacles, the magnification is less than one. For example in the present case, a –8.00-D myopic eye corrected in the 16-mm spectacle plane would get around 15% magnification after pIOL surgery. Thus, if no other effects were involved, visual acuity should improve about 2-3 lines after surgery for myopia, and the effect will obviously increase with the magnitude of the correction. This increase in visual acuity agrees with the change in visual acuity between before and after surgery reported for both eyes in the present patient. Similar effects would be found considering laser corneal refractive surgery, but corneal refractive surgery increases optical aberrations of the eye as well, therefore the acuity increase would be smaller than that predicted by magnification alone, and better outcomes might be expected when a pIOL is implanted. Small residual refractive errors occurring after pIOL implantation may be corrected using laser corneal surgery, which has been proved to be safe and effective combined with pIOLs [7]. No sight-threatening and visual complications were observed in the present study: high IOP, halos, glare, inflammation (e.g. acute uveitis), cataracts or pigment deposits, for example. One of the main problems with anterior chamber pIOL implantation is the potential for damage to the anterior segment structures, especially the corneal endothelium. However, previous studies indicate that new models of angle-supported and iris-fixated pIOLs are well tolerated by the corneal endothelium as improvement in IOL design and quality of surgery is achieved [8]. In addition, as the Artiflex pIOL has no angular support, the possibility of peripheral contact with the corneal endothelium is diminished. Unlike posterior chamber pIOLs, the Artiflex IOL is located far from the crystalline lens, thus decreasing the possibility of cataract complication. One interesting factor to consider in contact lens and surgical procedures (laser corneal refractive surgery or pIOL implantation) is the fact that the patient is looking along the visual axis of his correction for a far greater proportion of the time than with spectacles, since the correction moves with the eye. This becomes particularly important for the patient with high refractive error. It may be suggested that pIOL implantation can diminish the amplitude and frequency of the nystagmus, considering that the mechanism for this reduction might be the additional vergence and accommodative effort generated by the pIOL. In the present case, the amplitude and frequency of the nystagmus was not changed after pIOL surgery. In addition, it is not expected that ocular movement due to the nystagmus will affect the stability of the pIOL in a short-medium period of time. A large sample and longer follow-up would be necessary, however. In conclusion, this case report shows that pIOL implantation can be a safe therapeutic option to correct high refractive errors in patients with nystagmus. Some of these patients may greatly benefit from the surgery, which may improve their visual acuity and quality of life.

Disclosure Statement

The authors indicate no financial support or financial conflict of interest.
  8 in total

Review 1.  Examination and refractive management of patients with nystagmus.

Authors:  R W Hertle
Journal:  Surv Ophthalmol       Date:  2000 Nov-Dec       Impact factor: 6.048

2.  Laser in situ keratomileusis in myopic patients with congenital nystagmus.

Authors:  Ori Mahler; Ami Hirsh; Israel Kremer; Irina S Barequet; Arie L Marcovich; Pinhas Nemet; Samuel Levinger
Journal:  J Cataract Refract Surg       Date:  2006-03       Impact factor: 3.351

3.  Angle-supported phakic intraocular lenses followed by laser-assisted in situ keratomileusis for the correction of high myopia.

Authors:  Gonzalo Muñoz; Jorge L Alió; Robert Montés-Micó; José I Belda
Journal:  Am J Ophthalmol       Date:  2003-09       Impact factor: 5.258

Review 4.  Infantile nystagmus: current concepts in diagnosis and management.

Authors:  Larry A Abel
Journal:  Clin Exp Optom       Date:  2006-03       Impact factor: 2.742

5.  Simplified system of marking the cornea for a toric intraocular lens.

Authors:  John M Graether
Journal:  J Cataract Refract Surg       Date:  2009-09       Impact factor: 3.351

Review 6.  Phakic intraocular lens implantation for the correction of myopia: a report by the American Academy of Ophthalmology.

Authors:  David Huang; Steven C Schallhorn; Alan Sugar; Ayad A Farjo; Parag A Majmudar; William B Trattler; David J Tanzer
Journal:  Ophthalmology       Date:  2009-11       Impact factor: 12.079

7.  Role of contact lenses in the management of congenital nystagmus.

Authors:  E D Allen; P D Davies
Journal:  Br J Ophthalmol       Date:  1983-12       Impact factor: 4.638

8.  Phakic anterior chamber lenses for the correction of myopia: a 7-year cumulative analysis of complications in 263 cases.

Authors:  J L Alió; F de la Hoz; J J Pérez-Santonja; J M Ruiz-Moreno; J A Quesada
Journal:  Ophthalmology       Date:  1999-03       Impact factor: 12.079

  8 in total
  1 in total

1.  Corneal Collagen Cross-Linking Combined with an Artiflex Iris-Fixated Anterior Chamber Phakic Intraocular Lens Implantation in a Patient with Progressive Keratoconus.

Authors:  Sharif Hashmani; Nauman Hashmani; Roha Saeed Memon
Journal:  Case Rep Ophthalmol       Date:  2017-10-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.