[New method of measuring effective energy using copper-pipe absorbers in X-ray CT].

Hiroji Iida1, Kimiya Noto, Wataru Mitsui, Tadanori Takata, Tomoyuki Yamamoto, Kousuke Matsubara.   

Abstract

The general method of measuring the half-value layer (HVL) for X-ray computed tomography (CT) using square aluminum-sheet filters is inconvenient in that the X-ray tube has to be set to stationary mode. To avoid this inconvenience, we investigated a new method using copper-pipe filters that cover the ionization chamber (copper-pipe method). Using this method, the HVL can be measured at the isocenter in the rotation modes of CT. We examined the accuracy and reproducibility of the copper-pipe method compared with those of the general method. The effective energy measured using the copper-pipe method correlated well with the general method (y=1.064x, r=0.987), and its error was 1.81±1.38%. The results indicate that the copper-pipe method enables accurate measurement of the effective energy of X-ray CT and is a convenient method suited to all general X-ray equipment as well as all X-ray CT.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21937842     DOI: 10.6009/jjrt.67.1183

Source DB:  PubMed          Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi        ISSN: 0369-4305


  4 in total

1.  Using 80 kVp on a 320-row scanner for hepatic multiphasic CT reduces the contrast dose by 50 % in patients at risk for contrast-induced nephropathy.

Authors:  Narumi Taguchi; Seitaro Oda; Daisuke Utsunomiya; Yoshinori Funama; Takeshi Nakaura; Masanori Imuta; Sadahiro Yamamura; Hideaki Yuki; Masafumi Kidoh; Kenichiro Hirata; Tomohiro Namimoto; Masahiro Hatemura; Noriyuki Kai; Yasuyuki Yamashita
Journal:  Eur Radiol       Date:  2016-05-30       Impact factor: 5.315

2.  Evaluation of the effectiveness of X-ray protective aprons in experimental and practical fields.

Authors:  Hiroshige Mori; Kichiro Koshida; Osamu Ishigamori; Kosuke Matsubara
Journal:  Radiol Phys Technol       Date:  2013-12-13

3.  Validation of Computed Tomography-based Attenuation Correction of Deviation between Theoretical and Actual Values in Four Computed Tomography Scanners.

Authors:  Nobuhiro Yada; Hideo Onishi
Journal:  Asia Ocean J Nucl Med Biol       Date:  2016

4.  Feasibility of the new copper pipe method for evaluating half-value layer in computed tomography: A measurement and Monte Carlo simulation study.

Authors:  Rena Okubo; Kosuke Matsubara; Thunyarat Chusin; Tomoya Hibino; Yusuke Ito
Journal:  J Appl Clin Med Phys       Date:  2019-11-25       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.