Literature DB >> 21934335

In vitro comparison of noise levels produced by different CPAP generators.

Lieselotte Kirchner1, Martin Wald, Valerie Jeitler, Arnold Pollak.   

Abstract

INTRODUCTION: Minimization of noise exposure is an important aim of modern neonatal intensive care medicine. Binasal continuous positive airway pressure (CPAP) generators are among the most important sources of continuous noise in neonatal wards. The aim of this study was to find out which CPAP generator creates the least noise.
METHOD: In an experimental setup, two jet CPAP generators (Infant Flow® generator and MediJet®) and two conventional CPAP generators (Bubble CPAP® and Baby Flow®) were compared. Noise production was measured in decibels in an A-weighted scale [dB(A)] in a closed incubator at 2 mm lateral distance from the end of the nasal prongs. Reproduction of constant airway pressure and air leak was achieved by closure of the nasal prongs with a type of adhesive tape that is semipermeable to air.
RESULTS: The noise levels produced by the four generators were significantly different (p < 0.001). Values measured at a continuous constant flow rate of 8 l/min averaged 83 dB(A) for the Infant Flow® generator with or without sound absorber, 72 dB(A) for the MediJet®, 62 dB(A) for the Bubble CPAP® and 55 dB(A) for the Baby Flow®.
CONCLUSION: Conventional CPAP generators work more quietly than the currently available jet CPAP generators.
Copyright © 2011 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2011        PMID: 21934335     DOI: 10.1159/000329558

Source DB:  PubMed          Journal:  Neonatology        ISSN: 1661-7800            Impact factor:   4.035


  2 in total

1.  A Novel Approach to Improving Fat Delivery in Neonatal Enteral Feeding.

Authors:  Jane Jarjour; Alexa M Juarez; Denizen K Kocak; Nathan J Liu; Mika M Tabata; Keli M Hawthorne; Renata F Ramos; Steven A Abrams
Journal:  Nutrients       Date:  2015-06-23       Impact factor: 5.717

2.  Investigating Noise Exposure to Newborn Infants From Respiratory Support: Methodological Considerations.

Authors:  Devika Singh; Gerhard Fusch
Journal:  Cureus       Date:  2021-11-08
  2 in total