High-quality 3D correction of ring and radiant artifacts in flat panel detector-based cone beam volume CT imaging.

Emran Mohammad Abu Anas1, Jae Gon Kim, Soo Yeol Lee, Md Kamrul Hasan.   

Abstract

The use of an x-ray flat panel detector is increasingly becoming popular in 3D cone beam volume CT machines. Due to the deficient semiconductor array manufacturing process, the cone beam projection data are often corrupted by different types of abnormalities, which cause severe ring and radiant artifacts in a cone beam reconstruction image, and as a result, the diagnostic image quality is degraded. In this paper, a novel technique is presented for the correction of error in the 2D cone beam projections due to abnormalities often observed in 2D x-ray flat panel detectors. Template images are derived from the responses of the detector pixels using their statistical properties and then an effective non-causal derivative-based detection algorithm in 2D space is presented for the detection of defective and mis-calibrated detector elements separately. An image inpainting-based 3D correction scheme is proposed for the estimation of responses of defective detector elements, and the responses of the mis-calibrated detector elements are corrected using the normalization technique. For real-time implementation, a simplification of the proposed off-line method is also suggested. Finally, the proposed algorithms are tested using different real cone beam volume CT images and the experimental results demonstrate that the proposed methods can effectively remove ring and radiant artifacts from cone beam volume CT images compared to other reported techniques in the literature.

Entities:  

Mesh:

Year:  2011        PMID: 21934193     DOI: 10.1088/0031-9155/56/19/020

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  A Ring Artifact Correction Method: Validation by Micro-CT Imaging with Flat-Panel Detectors and a 2D Photon-Counting Detector.

Authors:  Mohamed Elsayed Eldib; Mohamed Hegazy; Yang Ji Mun; Myung Hye Cho; Min Hyoung Cho; Soo Yeol Lee
Journal:  Sensors (Basel)       Date:  2017-01-30       Impact factor: 3.576

2.  Complete Ring Artifacts Reduction Procedure for Lab-Based X-ray Nano CT Systems.

Authors:  Jakub Šalplachta; Tomáš Zikmund; Marek Zemek; Adam Břínek; Yoshihiro Takeda; Kazuhiko Omote; Jozef Kaiser
Journal:  Sensors (Basel)       Date:  2021-01-01       Impact factor: 3.576