Optimized knot placement for B-splines in deformable image registration.

Travis J Jacobson1, Martin J Murphy.   

Abstract

PURPOSE: To develop an automatic knot placement algorithm to enable the use of NonUniform Rational B-Splines (NURBS) in deformable image registration.
METHODS: The authors developed a two-step approach to fit a known displacement vector field (DVF). An initial fit was made with uniform knot spacing. The error generated by this fit was then assigned as an attractive force pulling on the knots, acting against a resistive spring force in an iterative equilibration scheme. To demonstrate the accuracy gain of knot optimization over uniform knot placement, we compared the sum of the squared errors and the frequency of large errors.
RESULTS: Fits were made to a one-dimensional DVF using 1-20 free knots. Given the same number of free knots, the optimized, nonuniform B-spline fit produced a smaller error than the uniform B-spline fit. The accuracy was improved by a mean factor of 4.02. The optimized B-spline was found to greatly reduce the number of errors more than 1 standard deviation from the mean error of the uniform fit. The uniform B-spline had 15 such errors, while the optimized B-spline had only two. The algorithm was extended to fit a two-dimensional DVF using control point grid sizes ranging from 8 x 8 to 15 x 15. Compared with uniform fits, the optimized B-spline fits were again found to reduce the sum of squared errors (mean ratio = 2.61) and number of large errors (mean ratio = 4.50).
CONCLUSIONS: Nonuniform B-splines offer an attractive alternative to uniform B-splines in modeling the DVF. They carry forward the mathematical compactness of B-splines while simultaneously introducing new degrees of freedom. The increased adaptability of knot placement gained from the generalization to NURBS offers increased local control as well as the ability to explicitly represent topological discontinuities.

Entities:  

Mesh:

Year:  2011        PMID: 21928630      PMCID: PMC3172863          DOI: 10.1118/1.3609416

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  The adaptive bases algorithm for intensity-based nonrigid image registration.

Authors:  Gustavo K Rohde; Akram Aldroubi; Benoit M Dawant
Journal:  IEEE Trans Med Imaging       Date:  2003-11       Impact factor: 10.048

2.  Fast parametric elastic image registration.

Authors:  Jan Kybic; Michael Unser
Journal:  IEEE Trans Image Process       Date:  2003       Impact factor: 10.856

3.  Objective assessment of deformable image registration in radiotherapy: a multi-institution study.

Authors:  Rojano Kashani; Martina Hub; James M Balter; Marc L Kessler; Lei Dong; Lifei Zhang; Lei Xing; Yaoqin Xie; David Hawkes; Julia A Schnabel; Jamie McClelland; Sarang Joshi; Quan Chen; Weiguo Lu
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

4.  4D-CT motion estimation using deformable image registration and 5D respiratory motion modeling.

Authors:  Deshan Yang; Wei Lu; Daniel A Low; Joseph O Deasy; Andrew J Hope; Issam El Naqa
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

5.  Evaluation of deformable registration of patient lung 4DCT with subanatomical region segmentations.

Authors:  Ziji Wu; Eike Rietzel; Vlad Boldea; David Sarrut; Gregory C Sharp
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

6.  Results of a multi-institution deformable registration accuracy study (MIDRAS).

Authors:  Kristy K Brock
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-11-10       Impact factor: 7.038

  6 in total
  5 in total

1.  Adaptive radiotherapy for NSCLC patients: utilizing the principle of energy conservation to evaluate dose mapping operations.

Authors:  Hualiang Zhong; Indrin J Chetty
Journal:  Phys Med Biol       Date:  2017-05-05       Impact factor: 3.609

2.  Segmented polynomials for incidence rate estimation from prevalence data.

Authors:  Severin Guy Mahiané; Oliver Laeyendecker
Journal:  Stat Med       Date:  2016-09-26       Impact factor: 2.373

3.  Individually wide range of renal motion evaluated by four-dimensional computed tomography.

Authors:  Hideomi Yamashita; Mami Yamashita; Masahiko Futaguchi; Ryousuke Takenaka; Shino Shibata; Kentaro Yamamoto; Akihiro Nomoto; Akira Sakumi; Satoshi Kida; Yoshihiro Kaneko; Shigeharu Takenaka; Takashi Shiraki; Keiichi Nakagawa
Journal:  Springerplus       Date:  2014-03-07

4.  From nonlinear optimization to convex optimization through firefly algorithm and indirect approach with applications to CAD/CAM.

Authors:  Akemi Gálvez; Andrés Iglesias
Journal:  ScientificWorldJournal       Date:  2013-11-24

5.  The effects of computed tomography image characteristics and knot spacing on the spatial accuracy of B-spline deformable image registration in the head and neck geometry.

Authors:  Charlotte L Brouwer; Roel G J Kierkels; Aart A van 't Veld; Nanna M Sijtsema; Harm Meertens
Journal:  Radiat Oncol       Date:  2014-07-29       Impact factor: 3.481

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.