Literature DB >> 21926639

Using continuous motion monitoring technology to determine patient's risk for development of pressure ulcers.

Eyal Zimlichman1, Zvika Shinar, Ronen Rozenblum, Shiraz Levkovich, Sarit Skiano, Martine Szyper-Kravitz, Arie Altman, Howard Amital, Yehuda Shoenfeld.   

Abstract

OBJECTIVES: To perform initial validation of a continuous motion monitoring technology that can potentially be used as a risk assessment tool to determine risk for developing pressure ulcers (PUs).
METHODS: We have used the EverOn system (Earlysense LTD, Ramat Gan, Israel) as a bed movement and activity monitor. The EverOn is a contactless continuous measurement system based on a piezoelectric sensor that is placed under the patient's mattress. The study was a noninterventional study performed in 2 medical departments in 2 medical centers. Recorded movement data from enrolled patients were retrospectively analyzed, and patients were assigned a motion level score. Motion scores for the first night of hospitalization were correlated with the Norton scale as calculated per patient on admission.
RESULTS: Overall, 116 patients were included in the study from the 2 sites. Motion score was significantly different between the PU risk groups as determined by the Norton scale (10.7 ± 6.2 for low, 5.4 ± 4.9 for intermediate, and 1.6 ± 3.2 for high risk; P < 0.001). Using the Norton scale as a gold standard to define high risk for developing PU (≤14), the sensitivity of the motion score was 85%, and the specificity was 93%. With regard to individual risk components, we found that activity, mobility, physical condition, and incontinence correlated highly with motion level.
CONCLUSIONS: The high correlation between the EverOn motion score and the calculated Norton scale indicates the potential of this technology to serve as a risk assessment tool for the development of PUs.

Entities:  

Mesh:

Year:  2011        PMID: 21926639     DOI: 10.1097/PTS.0b013e318230e6ef

Source DB:  PubMed          Journal:  J Patient Saf        ISSN: 1549-8417            Impact factor:   2.844


  3 in total

1.  Validation of Contact-Free Sleep Monitoring Device with Comparison to Polysomnography.

Authors:  Asher Tal; Zvika Shinar; David Shaki; Shlomi Codish; Aviv Goldbart
Journal:  J Clin Sleep Med       Date:  2017-03-15       Impact factor: 4.062

Review 2.  Monitoring healthy and disturbed sleep through smartphone applications: a review of experimental evidence.

Authors:  Edita Fino; Michela Mazzetti
Journal:  Sleep Breath       Date:  2018-04-23       Impact factor: 2.816

3.  Home monitoring of heart failure patients at risk for hospital readmission using a novel under-the-mattress piezoelectric sensor: A preliminary single centre experience.

Authors:  Mosi K Bennett; Mingyuan Shao; Eiran Z Gorodeski
Journal:  J Telemed Telecare       Date:  2016-07-09       Impact factor: 6.184