Literature DB >> 21925921

Concave polyethylene component improves biomechanical performance in lumbar total disc replacement--modified compressive-shearing test by finite element analysis.

Wen-Chuan Chen1, Yu-Liang Liu, Kun-Jhih Lin, Colin Joseph McClean, Hung-Jen Lai, Chi-Wei Chou, Tsung-Wei Chang, Chan-Tsung Yang, Chang-Hung Huang, Yu-Shu Lai, Cheng-Kung Cheng.   

Abstract

Failure of ultra-high molecular weight polyethylene components after total disc replacements in the lumbar spine has been reported in several retrieval studies, but immediate biomechanical evidence for those mechanical failures remained unclear. Current study aimed to investigate the failure mechanisms of commercial lumbar disc prostheses and to enhance the biomechanical performances of polyethylene components by modifying the articulating surface into a convex geometry. Modified compressive-shearing tests were utilized in finite element analyses for comparing the contact, tensile, and shearing stresses on two commercial disc prostheses and on a concave polyethylene design. The influence of radial clearance on stress distributions and prosthetic stability were considered. The modified compressive-shearing test revealed the possible mechanisms for transverse and radial cracks of polyethylene components, and would be helpful in observing the mechanical risks in the early design stage. Additionally, the concave polyethylene component exhibited lower contact and shearing stresses and more acceptable implant stability when compared with the convex polyethylene design through all radial clearances. Use of a concave polyethylene component in lumbar disc replacements decreased the risk of transverse and radial cracks, and also helped to maintain adequate stability. This design concept should be considered in lumbar disc implant designs in the future.
Copyright © 2011 IPEM. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21925921     DOI: 10.1016/j.medengphy.2011.08.010

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  2 in total

1.  We Need to Talk about Lumbar Total Disc Replacement.

Authors:  Stephen Beatty
Journal:  Int J Spine Surg       Date:  2018-08-03

2.  ISASS Policy Statement - Lumbar Artificial Disc.

Authors:  Jack Zigler; Rolando Garcia
Journal:  Int J Spine Surg       Date:  2015-03-12
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.