H J Zweifel1, H F Reinhardt, G A Horstmann, E Amrein.
Abstract
Today the accuracy of medical imaging systems lies between +/- 0.5 mm (CT) and +/- 2.5 mm (MRI). In neurosurgery the overall uncertainty may reach the dimension centimetres. A non-tactile ultrasonic measurement system is described that allows three-dimensional measurement with a theoretical accuracy of +/- 0.1 mm in a frequency range between 50 and 70 kHz. In practice the accuracy is better than +/- 1 mm with a repetition rate of 10 per second for the x,y,z-data sets.Mesh:
Year: 1990 PMID: 2192451 DOI: 10.1055/s-2007-1011533
Source DB: PubMed Journal: Ultraschall Med ISSN: 0172-4614 Impact factor: 6.548