Wear and surface analysis of 38 mm ceramic-on-metal total hip replacements under standard and severe wear testing conditions.

S R Williams1, J J Wu, A Unsworth, I Khan.   

Abstract

The purpose of this study was to compare the wear of zirconia-toughened alumina (ZTA) and alumina femoral heads tested against as-cast CoCrMo alloy acetabular cups under both standard and severe wear conditions. A new severe test, which included medio-lateral displacement of the head and rim impact upon relocation, was developed. This resulted in an area of metal transfer and an area of increased wear on the superior-anterior segment of the head that were thought to be due to dislocation and rim impact respectively. While the wear of all ceramic heads was immeasurable using the gravimetric method, the wear rates for the metallic cups from each test were readily calculated. An average steady state wear rate of 0.023 +/- 0.005 mm3/10(6) cycles was found for the cups articulating against ZTA under standard wear conditions. A similar result had previously been obtained for the wear of cups articulated against alumina heads of the same size (within the same laboratory). Under severe wear conditions an increase in the metallic cup steady state wear rate was found with the ZTA and alumina tests giving 0.623 +/- 0.252 and 1.35 +/- 0.154 mm3/10(6) cycles respectively. Wear of the ceramic heads was detected using atomic force microscopy which showed, under severe wear conditions, a decrease in polishing marks and occasional grain removal. The surfaces of the ZTA heads tested under standard conditions were virtually unchanged from the unworn samples. Friction tests showed low friction factors for all components, pre and post wear.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21922955     DOI: 10.1177/0954411911404773

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  4 in total

1.  Biotribological study of large diameter ceramic-on-CFR-PEEK hip joint including fluid uptake, wear and frictional heating.

Authors:  Qian Qian Wang; Jun Jie Wu; Anthony Unsworth; Adam Briscoe; Marcus Jarman-Smith; Conor Lowry; David Simpson; Simon Collins
Journal:  J Mater Sci Mater Med       Date:  2012-03-28       Impact factor: 3.896

2.  No Difference in Reoperations at 2 Years Between Ceramic-on-metal and Metal-on-metal THA: A Randomized Trial.

Authors:  C Anderson Engh; Supatra Sritulanondha; Abigail Korczak; Terrence David Whalen; Douglas D R Naudie; Richard W McCalden; Steven J MacDonald
Journal:  Clin Orthop Relat Res       Date:  2016-02       Impact factor: 4.176

3.  Wear performance of ceramic-on-metal hip bearings.

Authors:  Jörn Reinders; Robert Sonntag; Christian Heisel; Tobias Reiner; Leo Vot; Jan Philippe Kretzer
Journal:  PLoS One       Date:  2013-08-29       Impact factor: 3.240

4.  Zirconia phase transformation in retrieved, wear simulated, and artificially aged ceramic femoral heads.

Authors:  Maria Parkes; Kathryn Sayer; Markus Goldhofer; Philippa Cann; William L Walter; Jonathan Jeffers
Journal:  J Orthop Res       Date:  2017-06-20       Impact factor: 3.494

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.