Effect of implant-abutment connection design on load bearing capacity and failure mode of implants.

Stephanie Dittmer1, Marc Philipp Dittmer, Philipp Kohorst, Michael Jendras, Lothar Borchers, Meike Stiesch.   

Abstract

PURPOSE: In this in vitro study, six implant-abutment connection designs were compared and evaluated regarding load bearing capacities and failure modes.
MATERIALS AND METHODS: Five implants of Astra Tech, Bego, Camlog, Friadent, Nobel Biocare, and Straumann were separately embedded in stainless steel tubes using polyurethane, for a total of 30 specimens. Specimens were statically loaded under an angle of 30° with respect to the implant axis in a universal testing machine using a test setup according to ISO 14801. Failure was indicated by a load drop of 100 N in force. Load-displacement curves were analyzed, and maximum force and force at which permanent deformation occurred were recorded. Statistical analysis was performed using one-way ANOVA with the level of significance set at 0.05.
RESULTS: Statistical analysis revealed that the type of implant-abutment connection design has a significant influence on load bearing capacity (p < 0.001). The mean maximum forces ranged between 606 N (Straumann) and 1129 N (Bego); the forces where plastic deformation set in ranged between 368 N (Friadent) and 955 N (Bego). Failure modes differed between the various implant-abutment connection types tested.
CONCLUSIONS: Implant-abutment connection design has a significant influence on load bearing capacity and failure mode of implants; however, all implant-abutment connection designs tested would be expected to withstand clinically relevant forces.
© 2011 by the American College of Prosthodontists.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21910778     DOI: 10.1111/j.1532-849X.2011.00758.x

Source DB:  PubMed          Journal:  J Prosthodont        ISSN: 1059-941X            Impact factor:   2.752


  10 in total

1.  Validation of finite-element simulations with synchrotron radiography - A descriptive study of micromechanics in two-piece dental implants.

Authors:  Wolfram Wiest; Alexander Rack; Simon Zabler; Alex Schaer; Michael Swain; Katja Nelson
Journal:  Heliyon       Date:  2018-02-08

2.  Volumetric Changes in Morse Taper Connections After Implant Placement in Dense Bone. In-Vitro Study.

Authors:  Georgios E Romanos; Rafael Arcesio Delgado-Ruiz; Ana I Nicolas-Silvente
Journal:  Materials (Basel)       Date:  2020-05-16       Impact factor: 3.623

3.  A three-dimensional finite element analysis of the influence of varying implant crest module designs on the stress distribution to the bone.

Authors:  Shweta Maruti Patil; Abhijit Suresh Deshpande; Rahul Ramesh Bhalerao; Suryakant Bhanudas Metkari; Prithviraj Maruti Patil
Journal:  Dent Res J (Isfahan)       Date:  2019 May-Jun

4.  Deformation of the Internal Connection of Narrow Implants after Insertion in Dense Bone: An in Vitro Study.

Authors:  Rafael Delgado-Ruiz; Ana Nicolas Silvente; Georgios Romanos
Journal:  Materials (Basel)       Date:  2019-06-06       Impact factor: 3.623

5.  A Comparative 3D Finite Element Computational Study of Three Connections.

Authors:  Davide Farronato; Mattia Manfredini; Andrea Stevanello; Veronica Campana; Lorenzo Azzi; Marco Farronato
Journal:  Materials (Basel)       Date:  2019-09-26       Impact factor: 3.623

6.  Influence of Connection Type and Platform Diameter on Titanium Dental Implants Fatigue: Non-Axial Loading Cyclic Test Analysis.

Authors:  Ana I Nicolas-Silvente; Eugenio Velasco-Ortega; Ivan Ortiz-Garcia; Alvaro Jimenez-Guerra; Loreto Monsalve-Guil; Raul Ayuso-Montero; Javier Gil; Jose Lopez-Lopez
Journal:  Int J Environ Res Public Health       Date:  2020-12-02       Impact factor: 3.390

7.  Rotational tolerances of a titanium abutment in the as-received condition and after screw tightening in a conical implant connection.

Authors:  Rosario Prisco; Giuseppe Troiano; Luigi Laino; Khrystyna Zhurakivska
Journal:  J Adv Prosthodont       Date:  2021-12-22       Impact factor: 1.904

8.  A Comparative Evaluation of the Strain Transmitted through Prostheses on Implants with Two Different Macro-Structures and Connection during Insertion and Loading Phase: An In Vitro Study.

Authors:  Riyam Kassem; Amar Samara; Ameer Biadsee; Shchada Masarwa; Tarek Mtanis; Zeev Ormianer
Journal:  Materials (Basel)       Date:  2022-07-16       Impact factor: 3.748

9.  In situ microradioscopy and microtomography of fatigue-loaded dental two-piece implants.

Authors:  Wolfram Wiest; Simon Zabler; Alexander Rack; Christian Fella; Andreas Balles; Katja Nelson; Rainer Schmelzeisen; Randolf Hanke
Journal:  J Synchrotron Radiat       Date:  2015-10-09       Impact factor: 2.616

10.  Load-Bearing Capacity and Retention of Newly Developed Micro-Locking Implant Prosthetic System: An In Vitro Pilot Study.

Authors:  Jae-Won Choi; Kyung-Hee Choi; Hee-Jin Chae; Sung-Ki Chae; Eun-Bin Bae; Jin-Ju Lee; So-Hyoun Lee; Chang-Mo Jeong; Jung-Bo Huh
Journal:  Materials (Basel)       Date:  2018-04-06       Impact factor: 3.623

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.