Abstract

Proper acetabular cup orientation is essential in total hip arthroplasty. The purpose of this study was to evaluate the accuracy of a particular imageless computer navigation system in determining cup position. Thirty-nine computer-navigated total hip arthroplasty intraoperative measurements of cup abduction and anteversion were compared with those from follow-up radiographs. Sensitivity, specificity, accuracy, prevalence-adjusted positive value (PPV), and negative predictive value were calculated for both navigation and radiographs. Navigation measurements had high specificity and PPV when assessing cup abduction and anteversion (specificity >90%, PPV >94%). In contrast, the system was not very effective in detecting suboptimal cup position (sensitivity abduction, 50%; anteversion, 33%). Intraoperative navigation readings in the safe zone have high probability of indicating correct placement. However, confirmation of suboptimal cup position intraoperatively requires additional diagnostic methods. Copyright Â
© 2012 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21908165     DOI: 10.1016/j.arth.2011.06.022

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  3 in total

1.  Anteroposterior Radiographs Are More Accurate than Cross-Table Lateral Radiographs for Acetabular Anteversion Assessment: a Retrospective Cohort Study.

Authors:  Peter K Sculco; Alexander S McLawhorn; Kaitlin M Carroll; Benjamin A McArthur; David J Mayman
Journal:  HSS J       Date:  2015-10-19

2.  Abnormal spinopelvic mobility as a risk factor for acetabular placement error in total hip arthroplasty using optical computer-assisted surgical navigation system.

Authors:  Seong J Jang; Jonathan M Vigdorchik; Eric W Windsor; Ran Schwarzkopf; David J Mayman; Peter K Sculco
Journal:  Bone Jt Open       Date:  2022-06

3.  Lewinnek Safe Zone References are Frequently Misquoted.

Authors:  Aonnicha Burapachaisri; Ameer Elbuluk; Edem Abotsi; Jim Pierrepont; Seth A Jerabek; Aaron J Buckland; Jonathan M Vigdorchik
Journal:  Arthroplast Today       Date:  2020-11-26
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.