Statistical control of the spectral quality index in electron beams.

José Manuel de la Vega1, Rafael Jesús Martínez-Luna, Damián Guirado, Manuel Vilches, Antonio Miguel Lallena.   

Abstract

BACKGROUND AND
PURPOSE: In radiotherapy with electron linear accelerators, it is mandatory to guarantee the stability of beam output spectra. In this work we study the quality control of the magnitude that provides the beam spectral quality, by following statistical techniques usual in industry process.
METHODS: We establish the differences between the measurement of the reference quality index, R(50,ion), which involves usually a costly procedure, and that of the index used for the statistical control, which can be done in an easier way and with a higher frequency. The methodology developed is applied to a Siemens Mevatron KDS electron linear accelerator, specifically for the 6, 12 and 18 MeV modes. The uncertainties of both procedures are evaluated and are correlated. A retrospective analysis of the data registered during more than 13 years is carried out using CUSUM algorithmic and moving range charts.
RESULTS: The larger uncertainties in R(50,ion) come from the measurement device and the observer. The uncertainty in the measurement of the statistical control variable, which is mainly due to the geometry setup, is negligible against its intrinsic variability. The application of CUSUM charts pointed out different out of control situations linked in much cases to assignable causes. In addition, moving range charts showed up situations of anomalous functioning of the LINAC.
CONCLUSIONS: Algorithmic CUSUM charts permit a nice monitoring of R(50,ion). If the variability is not controlled, by using charts such as the moving range ones, the number of false alarms could increase without a change in the mean value of R(50,ion). Published by Elsevier Ireland Ltd.

Mesh:

Year:  2011        PMID: 21903284     DOI: 10.1016/j.radonc.2011.08.010

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  2 in total

1.  Evaluation of Elekta Agility multi-leaf collimator performance using statistical process control tools.

Authors:  Sandra M Meyers; Michael J Balderson; Daniel Létourneau
Journal:  J Appl Clin Med Phys       Date:  2019-06-14       Impact factor: 2.102

2.  Electron beam energy QA - a note on measurement tolerances.

Authors:  Juergen Meyer; Matthew J Nyflot; Wade P Smith; Landon S Wottoon; Lori Young; Fei Yang; Minsun Kim; Kristi R G Hendrickson; Eric Ford; Alan M Kalet; Ning Cao; Claire Dempsey; George A Sandison
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102