Comparison of non-contact methods for the measurement of central corneal thickness.

Ebru Gorgun1, Nursal Melda Yenerel, Umut Asli Dinc, Banu Oncel, Raciha Beril Kucumen, Deniz Oral, Ferda Ciftci.   

Abstract

BACKGROUND AND
OBJECTIVE: This study examined the repeatability of and agreements between central corneal thickness measurements obtained by four different non-contact pachymetry devices. PATIENTS AND METHODS: Seventy-eight eyes of 39 subjects were included. Central corneal thickness of each eye was measured by Visante optical coherence tomography (OCT) (Carl Zeiss Meditec Inc., Dublin, CA), Pentacam (Oculus Optikgeräte GmbH, Wetzlar, Germany), Orbscan IIz topography (Bausch & Lomb Surgical Inc., San Dimas, CA), and slit-lamp OCT (SL-OCT) (Heidelberg Engineering GmbH, Heidelberg, Germany). Inter-device agreements and correlations and repeatability of each device were examined.
RESULTS: All measurement methods correlated well with each other with a correlation coefficient greater than 0.90 and P value of less than .001 for all comparisons. However, Pentacam overestimated central corneal thickness: 546.7 ± 38.2, 535.5 ± 42.7, 531.7 ± 37.6, and 531.2 ± 36.0 μm for Pentacam, Orbscan IIz, Visante OCT, and SL-OCT, respectively (P < .001 for all comparisons versus Pentacam). Despite good correlation, magnitude of differences was high and this bias was proportional (ie, not constant across a range of corneal thickness values) for the following pairs: Orbscan versus Visante OCT, Orbscan versus SL-OCT, and Orbscan versus Pentacam (P < .001 for all comparisons).
CONCLUSION: Although measurements obtained by various non-contact methods correlate well, numerical agreement of the results may not be sufficient for their interchangeable use in clinical practice. Copyright 2011, SLACK Incorporated.

Entities:  

Mesh:

Year:  2011        PMID: 21899245     DOI: 10.3928/15428877-20110812-02

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging        ISSN: 1542-8877


  2 in total

1.  Central Corneal Thickness in Highly Myopic Eyes: Inter-device Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy.

Authors:  Nader Nassiri; Kourosh Sheibani; Sare Safi; Saman Nassiri; Alireza Ziaei; Alireza Ziaee; Farnaz Haji; Shiva Mehravaran; Nariman Nassiri
Journal:  J Ophthalmic Vis Res       Date:  2014-01

2.  Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology.

Authors:  Erhan Ozyol; Pelin Özyol
Journal:  Indian J Ophthalmol       Date:  2017-06       Impact factor: 1.848

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.