D Maser1, S Pandey, H Ring, M P Ledbetter, S Knappe, J Kitching, D Budker.
Abstract
We present magnetic detection of a single, 2 μm diameter cobalt microparticle using an atomic magnetometer based on a microfabricated vapor cell. These results represent an improvement by a factor of 10(5) in terms of the detected magnetic moment over previous work using atomic magnetometers to detect magnetic microparticles. The improved sensitivity is due largely to the use of small vapor cells. In an optimized setup, we predict detection limits of 0.17 μm(3).Entities:
Mesh:
Substances:
Year: 2011 PMID: 21895290 DOI: 10.1063/1.3626505
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523