Prediction of implantable pulse generator longevity in deep brain stimulation: limitations and possible solutions in clinical practice.

Colin D M Stewart1, Sam Eljamel.   

Abstract

BACKGROUND: Deep brain stimulation (DBS) is used in the treatment of movement disorders, chronic pain and certain neuropsychiatric conditions. Its effect is achieved through local stimulation of specific targets of the brain. Electrical pulses are delivered at settings that can be tailored to provide optimum symptom control. DBS is powered by an implantable pulse generator (IPG). These IPGs do not have an indefinite lifespan and will eventually become depleted. Therefore, regular checks are carried out to monitor the IPG's power status.
OBJECTIVE: To evaluate the utility of computer-based prediction of IPG longevity in patients with DBS devices by imputing the device's settings. PATIENTS: Eighty-two IPGs in 47 consecutive patients were studied comparing the actual longevity of the IPGs with their predicted longevity using a computer-based calculator.
RESULTS: Our study showed that the predicted longevities were too different from actual longevities to have a role in clinical practice. The time difference varied by more than 12 months in some cases, with a correlation coefficient of <0.68. There were multiple reasons for this.
CONCLUSION: The longevities of IPGs could not be predicted accurately enough for the computer-based calculator to have a role in clinical practice or the proactive scheduling of IPG replacement. However, rechargeable IPGs may help avoid premature replacements and battery depletions.
Copyright © 2011 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2011        PMID: 21894059     DOI: 10.1159/000329360

Source DB:  PubMed          Journal:  Stereotact Funct Neurosurg        ISSN: 1011-6125            Impact factor:   1.875


  3 in total

Review 1.  Deep brain stimulation (DBS) at the interface of neurology and psychiatry.

Authors:  Nolan R Williams; Michael S Okun
Journal:  J Clin Invest       Date:  2013-11-01       Impact factor: 14.808

Review 2.  Disease-specific longevity of impulse generators in deep brain stimulation and review of the literature.

Authors:  Christoph van Riesen; Georg Tsironis; Doreen Gruber; Fabian Klostermann; Patricia Krause; Gerd Helge Schneider; Andreas Kupsch
Journal:  J Neural Transm (Vienna)       Date:  2016-05-19       Impact factor: 3.575

3.  Clinical and programming pattern of patients with impending deep brain stimulation power failure: a retrospective chart review.

Authors:  Raja Mehanna; Kathy M Wilson; Scott E Cooper; Andre G Machado; Hubert H Fernandez
Journal:  J Clin Mov Disord       Date:  2014-11-20