Intraaortic balloon pump timing discrepancies in adult patients.

George M Pantalos1, Steven C Koenig, Kevin J Gillars, Gilbert S Haugh, Robert D Dowling, Laman A Gray.   

Abstract

The objective of this clinical study was to quantify the incidence and magnitude of intraaortic balloon pump (IABP) inflation and deflation landmark discrepancies associated with the IABP catheter arterial pressure waveform. Cardiac surgery patients with an IABP inserted prior to surgery were recruited. Following cardiac exposure, a high-fidelity pressure catheter was inserted into the aortic root for digital recording. The radial artery pressure signal was simultaneously recorded from the patient monitor along with the arterial pressure and electrocardiogram waveforms from the IABP console while operating at 1:1 and 1:2 synchronization. In selected patients, recordings were obtained with the IABP timed to the high-fidelity aortic root waveform. In all 11 patients, inflation and deflation landmark delays were observed when comparing the aortic root waveforms to the IABP arterial pressure waveforms (inflation delay = 74 ± 29 [23-117] ms; deflation delay = 71 ± 37 [24-141] ms, mean ± standard deviation [min-max]). Delays were greater when compared to the radial artery waveform (inflation delay = 175 ± 50 [100-233] ms; deflation delay = 168 ± 52 [100-274] ms). In all cases, the landmark delays were statistically different from zero (P < 0.001). Diastolic augmentation and afterload reduction varied with waveform source. Conflicting indications of afterload reduction occurred in four patients. Timing to the aortic root waveform resulted in greater diastolic pressure augmentation and afterload reduction but mixed changes in stroke volume. Delay and distortion of the arterial waveform was consistently found when measured through the IABP catheter lumen. These delays can alter IABP efficacy and may be eliminated by using high-fidelity sensing of aortic pressure.
© 2011, Copyright the Authors. Artificial Organs © 2011, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21883317     DOI: 10.1111/j.1525-1594.2011.01292.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  2 in total

1.  End-diastolic flow reversal limits the efficacy of pediatric intra-aortic balloon pump counterpulsation.

Authors:  Carlo R Bartoli; Benjamin D Rogers; Constantine E Ionan; George M Pantalos
Journal:  J Thorac Cardiovasc Surg       Date:  2013-10-15       Impact factor: 5.209

2.  Determination of cardiac output from pulse pressure contour during intra-aortic balloon pumping in patients with low ejection fraction.

Authors:  Jos R C Jansen; Marcelo B Bastos; Pat Hanlon; Nicolas M Van Mieghem; Ottavio Alfieri; Jan J Schreuder
Journal:  J Clin Monit Comput       Date:  2019-05-14       Impact factor: 2.502

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.