Literature DB >> 21883036

An assessment of dose accuracy and injection force of a novel prefilled insulin pen: comparison with a widely used prefilled insulin pen.

Jakob O Wielandt1, Marcus Niemeyer, Marianne R Hansen, Ditlef Bucher, Niels B Thomsen.   

Abstract

OBJECTIVE: FlexTouch® (FT) (Novo Nordisk A/S, Bagsværd, Denmark) is a new prefilled insulin pen with a novel injection mechanism encompassing no push-button extension at any dose-setting. This study assessed the dose accuracy and injection force of FT compared with the established Next Generation FlexPen® (NGFP) (Novo Nordisk A/S Bagsværd, Denmark). RESEARCH DESIGN AND METHODS: Dose accuracy was measured at the minimum, medium and maximum doses (FT, 1, 40 and 80 international units (IU) and NGFP, 1, 30 and 60 IU). Injection force was measured during the injection of the maximum dose (FT, 80 IU; NGFP, 60 IU). MAIN OUTCOMES: FT and NGFP delivered insulin accurately and consistently at all doses (mean ± s.d., FT at 1 IU, 0.98 ± 0.07; 40 IU, 39.86 ± 0.33; 80 IU, 79.76 ± 0.64; NGFP at 1 IU, 1.02 ± 0.08; 30 IU, 29.69 ± 0.30; 60 IU, 59.50 ± 0.51). FT had a significantly (p < 0.0001) lower injection force than NGFP.
CONCLUSIONS: The study demonstrated that FT and NGFP deliver insulin accurately and consistently at low, medium and high doses. The novel torque spring injection mechanism of FT results in a significantly lower injection force than NGFP and a pen requiring less thumb-pressure to inject insulin may be welcomed by many people with diabetes.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21883036     DOI: 10.1517/17425247.2011.615308

Source DB:  PubMed          Journal:  Expert Opin Drug Deliv        ISSN: 1742-5247            Impact factor:   6.648


  5 in total

1.  Patients' Preferences for Insulin Injection Devices.

Authors:  Andreas Pfützner
Journal:  J Diabetes Sci Technol       Date:  2017-01-09

2.  Safeguarding the process of drug administration with an emphasis on electronic support tools.

Authors:  Hanna M Seidling; Anette Lampert; Kristina Lohmann; Julia T Schiele; Alexander J F Send; Diana Witticke; Walter E Haefeli
Journal:  Br J Clin Pharmacol       Date:  2013-09       Impact factor: 4.335

3.  Taste and/or Odour Disturbances in Pediatric Patients Undergoing IV Flush with Normal Saline Administered by Prefilled Syringe.

Authors:  Steven J Celetti; Régis Vaillancourt; Elena Pascuet; Diane Sharp
Journal:  Can J Hosp Pharm       Date:  2012-09

4.  Model study of the pressure build-up during subcutaneous injection.

Authors:  Maria Thomsen; Anier Hernandez-Garcia; Joachim Mathiesen; Mette Poulsen; Dan N Sørensen; Lise Tarnow; Robert Feidenhans'l
Journal:  PLoS One       Date:  2014-08-14       Impact factor: 3.240

5.  Efficacy and safety of insulin degludec/insulin aspart versus biphasic insulin aspart 30 in Chinese adults with type 2 diabetes: A phase III, open-label, 2:1 randomized, treat-to-target trial.

Authors:  Wenying Yang; Jianhua Ma; Tianpei Hong; Ming Liu; Heng Miao; Yongde Peng; Changjiang Wang; Xiangjin Xu; Tao Yang; Anne M Nielsen; Lili Pan; Weihong Liu; Weigang Zhao
Journal:  Diabetes Obes Metab       Date:  2019-04-04       Impact factor: 6.577

  5 in total