Status of serum-calcium and -albumin measurement in Argentina assessed in 300 representative laboratories with 20 fresh frozen single donation sera.

Hedwig C M Stepman1, Dietmar Stöckl, Rosana Acheme, Sandra Sesini, Daniel Mazziotta, Linda M Thienpont.   

Abstract

BACKGROUND: The Fundación Bioquímica Argentina (FBA) performs external quality assessment (EQA) of >3200 laboratories. However, FBA realizes that sample non-commutability and predominant use of heterogeneous systems may bias the estimated performance and standardization status. To eliminate these confounding factors, a study using frozen single donation sera was undertaken with the focus on serum-calcium and -albumin measurement.
METHODS: Target values were established from the results produced with homogeneous systems. In groups of n=7, system effects were investigated. Laboratory performance was evaluated from the correlation coefficient r between the measurement results for all sera and the target values. This allowed ranking of the laboratories and judgment of the deviation for individual samples (total error) against a 10% limit. The total error specification was a deviation for ≥ 5 samples exceeding 10% and/or causing a result outside the laboratory's reference interval.
RESULTS: For calcium (n=303) (range: 2.06-2.42 mmol/L), 81 laboratories had an r-value <0.6, 43 even <0.4; the total error was relevant for 97 (10% limit) and 111 (reference interval) laboratories. For albumin (n=311) (range: 34.7-45.7 g/L) r was <0.7 (<0.4) in 44 (16) laboratories; 83 and 36 laboratories exceeded the total error criteria. Laboratories using homogeneous systems were generally ranked higher by correlation. System effects were moderate for calcium, but significant for albumin.
CONCLUSIONS: The study demonstrated the need to improve the quality and harmonization of calcium and albumin testing in the investigated laboratories. To achieve this objective, we promote co-operation between laboratories, EQA provider and manufacturers.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21875398     DOI: 10.1515/CCLM.2011.681

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694