Bong Joo Kang1, Sung Hun Kim, Byung Gil Choi.
Abstract
The object of this study was to compare of full-field digital mammography (FFDM) workstation and conventional picture archiving and communication systems (PACS) in image quality and diagnostic performance. We assembled 80 masses and 80 microcalcifications. Images were displayed on workstation, 5M, and 3M PACS monitors. The image quality for mammograms on workstation was significantly better than that for mammograms on PACS monitors. The sensitivity and NPV for microcalcifications on workstation were higher than those on PACS monitors. The conventional PACS cannot substitute for a FFDM workstation for mammographic evaluation.Mesh:
Year: 2011 PMID: 21872121 DOI: 10.1016/j.clinimag.2010.10.005
Source DB: PubMed Journal: Clin Imaging ISSN: 0899-7071 Impact factor: 1.605