C Ferreiro-Vera1, J M Mata-Granados, J M Quesada Gómez, M D Luque de Castro.
Abstract
The automated method developed for the determination of carotenoids uses 200 μL of serum, which was mixed with 400 μL of tetrahydrofuran, vortexed for 1 min, settled for 10 min, centrifuged for 6 min and the supernatant injected into an automatic solid-phase extraction (SPE) system for cleanup-preconcentration. A 10% water-acetonitrile mobile phase at 1.5 mL min(-1) eluted the retained compounds and transferred them on-line to a reversed-phase analytical column for individual separation of the target analytes. Visible detection was performed at 450 and 460 nm. The detection limits for the target analytes were between 3 and 30 ng mL(-1); the precision (expressed as relative standard deviation) ranged between 2.83 and 5.06% for repeatability and between 3.80 and 7.40% for within laboratory reproducibility. The total analysis time was 18 min. The proposed method is reliable, robust, and has an excellent potential for high-throughput use in both clinical and research laboratories.Entities: