Literature DB >> 2187032

A cytofluorometric method to quantify membrane antigens on individual alveolar macrophages.

W Pankow1, K Neumann, J Rüschoff, J Heymanns, P Von Wichert.   

Abstract

An immunofluorescence method for assaying membrane-bound antigens on individual alveolar macrophages (AM) collected by bronchoalveolar lavage (BAL) is described. Cells were labelled with FITC-conjugated anti-HLA-DR antibodies in a single (direct) step. Quantification of the fluorescence was performed by computer-assisted cytophotometry. Alveolar macrophages, especially when obtained from cigarette smokers, exhibited an autofluorescence which interfered with the measurement of specific fluorescence. The specific fluorescence was calculated by determination of total fluorescence in the wave-length optimal for FITC and the non-specific fluorescence in a different wavelength during a second measurement. Specificity and reproducibility testing confirmed the reliability of the method.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2187032     DOI: 10.1016/0022-1759(90)90429-y

Source DB:  PubMed          Journal:  J Immunol Methods        ISSN: 0022-1759            Impact factor:   2.303


  2 in total

1.  Reduction in HLA-DR antigen density on alveolar macrophages of smokers.

Authors:  W Pankow; K Neumann; J Rüschoff; R Schröder; P von Wichert
Journal:  Lung       Date:  1991       Impact factor: 2.584

2.  An immunohistochemical method for the detection of tumour necrosis factor alpha in cytospins of human bronchoalveolar lavage cells.

Authors:  S Hamilton; M Healy; P Corris; R Scott
Journal:  Histochem J       Date:  1995-06
  2 in total