A prefilled insulin pen with a novel injection mechanism and a lower injection force than other prefilled insulin pens.

Hans Hemmingsen1, Marcus Niemeyer, Marianne Rye Hansen, Ditlef Bucher, Niels Bjerrum Thomsen.   

Abstract

BACKGROUND: FlexTouch(®) (Novo Nordisk A/S, Bagsvaerd, Denmark) is a new prefilled insulin pen for people with diabetes, with a novel injection mechanism and no push-button extension at any dose setting. This study compared the injection force of FlexTouch with that of SoloStar(®) (sanofi-aventis, Paris, France) and KwikPen(®) (Eli Lilly & Co., Indianapolis, IN).
METHODS: Injection force was measured with the manufacturers' recommended needle attached to each pen (NovoFine(®) [Novo Nordisk] 32-gauge tip extra thin wall 6 mm needle for FlexTouch and BD [Franklin Lakes, NJ] MicroFine™ 31-gauge 5 mm needle for SoloStar and KwikPen) during injection of the maximum dose (60 IU for KwikPen and 80 IU for FlexTouch and SoloStar). Injection was performed at three different constant push-button speeds.
RESULTS: FlexTouch had a significantly (P<0.0001) lower injection force than SoloStar and KwikPen at all injection speeds. The mean±SD injection force of FlexTouch was 5.1±0.5 N. At 4, 6, and 8 mm/s push-button speeds, the injection force of SoloStar was 13.5±2.1, 19.1±1.9, and 26.9±2.4 N, respectively, and the injection force of KwikPen was 14.5±1.9, 20.9±1.4, and 28.2±1.4 N, respectively.
CONCLUSIONS: The injection mechanism of FlexTouch means that insulin injection is driven by a torque spring and not the thumb pressure of the user. This results in a 62-82% lower injection force with FlexTouch than other prefilled insulin pens.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21864019     DOI: 10.1089/dia.2011.0110

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  5 in total

1.  FlexTouch: An Insulin Pen-Injector with a Low Activation Force Across Different Insulin Formulations, Needle Technologies, and Temperature Conditions.

Authors:  Niels Gudiksen; Thibaud Hofstätter; Birgitte B Rønn; Thomas Sparre
Journal:  Diabetes Technol Ther       Date:  2017-07-13       Impact factor: 6.118

2.  Overview of insulin and non-insulin delivery devices in the treatment of diabetes.

Authors:  Michele Pisano
Journal:  P T       Date:  2014-12

3.  The value of fast-acting insulin aspart compared with insulin aspart for patients with diabetes mellitus treated with bolus insulin from a UK health care system perspective.

Authors:  Lalantha Leelarathna; Donna Ashley; Carrie Fidler; Witesh Parekh
Journal:  Ther Adv Endocrinol Metab       Date:  2018-04-04       Impact factor: 3.565

4.  A comparison study of patient ratings and safety of 32- and 34-gauge insulin pen needles.

Authors:  Satoru Yamada; Yoshifumi Yamada; Yoko Tsukamoto; Mitsuhisa Tabata; Junichiro Irie
Journal:  Diabetol Int       Date:  2015-10-28

5.  Non-stick syringe needles: Beneficial effects of thin film metallic glass coating.

Authors:  Jinn P Chu; Chia-Chi Yu; Yusuke Tanatsugu; Mikito Yasuzawa; Yu-Lin Shen
Journal:  Sci Rep       Date:  2016-08-30       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.