Literature DB >> 21861750

Impact of stent-graft oversizing on the thoracic aorta: experimental study in a porcine model.

Igor Rafael Sincos1, Ricardo Aun, Erasmo Simão da Silva, Sergio Belczak, Maria de Lourdes Higuchi, Vitor Cervantes Gornati, Pedro Nogueira Gigglio, Anna Paula Weinhardt Baptista, Luiz Francisco Poli de Figueiredo.   

Abstract

PURPOSE: To analyze in an experimental animal model the effect of 4 different levels of stents-graft oversizing on non-atherosclerotic aortas such as those found in young individuals who undergo stent-graft repair for traumatic aortic injuries.
METHODS: The diameter of the porcine thoracic aorta is similar to the aorta of young adults (18-20 mm), so 25 pigs were randomized into 5 groups: 1 control (without stent-graft) and 4 oversizing groups (A: 10%-19%, B: 20%-29%, C: 30%-39%, and D: >40%). Two types of biomechanical tests were performed on all aortas 4 weeks after endoprosthesis deployment.
RESULTS: The results of the detachment test, which analyzed the strength necessary to remove the stent-graft from the aorta, were similar in the 4 groups (A: 42 N, B: 41 N, C: 46 N, and D: 46 N). However, 2 aortas ruptured during the tests (groups C and D). The second test was performed in 3 aortic segments. Maximum shear strength, maximum stress, and maximum tension supported by the aortic wall had a negative and linear correlation with oversizing. There were significant differences in all 4 groups when compared with the control group. Strain, which reflects the elastic properties of the aortic wall, was very similar in all 4 groups, but a great difference was found when compared with the control group (p<0.0001).
CONCLUSION: The study showed an important subacute change in the biomechanical properties of the aortic wall after implantation of an oversized endoprosthesis. This weakness of the aortic wall was confirmed by 2 ruptures during the detachment test. These results partially explain the interaction of stent-grafts with non-atherosclerotic thoracic aortas and may serve as a basis for further studies and the development of specific material to be used in vascular trauma and young patients.

Entities:  

Mesh:

Year:  2011        PMID: 21861750     DOI: 10.1583/11-3470.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  4 in total

1.  Early histological changes in the porcine aortic media after thoracic stent-graft implantation.

Authors:  Johannes Scheumann; Claudia Heilmann; Friedhelm Beyersdorf; Matthias Siepe; Robert M Brenner; Dittmar Böckler; Randall B Griepp; Moritz S Bischoff
Journal:  J Endovasc Ther       Date:  2012-06       Impact factor: 3.487

2.  Multilayer stents affect the final diameter of aortic aneurysms and maintain renal artery patency for a short time in a swine experimental model.

Authors:  Anna Paula Weinhardt Baptista-Strazzi; Ricardo Aun; Igor Rafael Sincos; Allana M Tobita; Maria Fernanda Cassino Portugal; Vitória Penido de Paula; Oskar Kaufmann; Nelson Wolosker
Journal:  Clinics (Sao Paulo)       Date:  2021-05-21       Impact factor: 2.365

3.  Appearance and rapid evolution of thoraco-abdominal intramural hematoma after TEVAR.

Authors:  Edoardo Cervi; Franco Nodari; Emanuele Botteri; Girolomina Mazzeo; Bonardelli Stefano
Journal:  JRSM Cardiovasc Dis       Date:  2017-05-22

4.  Acute aortic wall injury caused by aortic cross-clamping: morphological and biomechanical study of the aorta in a swine model of three aortic surgery approaches.

Authors:  Marcela Polachini Prata; Rodrigo Gibin Jaldin; Pedro Luiz Toledo de Arruda Lourenção; Marcone Lima Sobreira; Ricardo de Alvarenga Yoshida; Simone Antunes Terra; Rosa Marlene Viero; Winston Bonetti Yoshida
Journal:  J Vasc Bras       Date:  2020-03-12
  4 in total