Literature DB >> 21861086

Histological findings of failed gold micro shunts in primary open-angle glaucoma.

Luca Agnifili1, Ciro Costagliola, Michele Figus, Giovanna Iezzi, Adriano Piattelli, Paolo Carpineto, Rodolfo Mastropasqua, Marco Nardi, Leonardo Mastropasqua.   

Abstract

BACKGROUND: To describe the histological features of failed gold micro shunts (GMS) in unsuccessful implantations for refractory primary open-angle glaucoma (POAG).
METHODS: This was an interventional case series study. Five eyes of five glaucomatous patients with unsuccessful GMS implantation underwent shunt removal. Each device was sectioned into three portions: proximal or anterior chamber (AC) portion, middle or scleral portion and distal or suprachoroidal (SC) portion. The histological analysis was performed throughout the whole extent of the shunt, describing both the inner spaces and the outer surface.
RESULTS: At the moment of removal all devices were correctly located into the SC space and in AC, with the exception of a case presenting corneal endothelial contact. The mean intra-ocular pressure before GMS removal was 30.4 ± 5.3 mmHg, and the mean time of GMS removal after implantation was 6.8 ± 2.5 months. No significant histological differences were documented among the five analyzed devices. The main feature was the presence of a thick connective capsule-like reaction surrounding both the proximal and distal ends and invading the posterior and anterior grid holes, whereas a more loosely arranged connective tissue was observed within the inner channels. Signs of surface fibrosis of the middle-scleral portion and inflammatory cell infiltration of the device were not documented in any of the cases.
CONCLUSIONS: Failed GMS implantations presented connective tissue filling all the inner spaces and creating a thick fibrotic capsule surrounding the ends of the device. This modification isolated the GMS from the AC and SC space, impeding aqueous flows throughout the shunt.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21861086     DOI: 10.1007/s00417-011-1778-6

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  27 in total

Review 1.  Recent advances in trabeculectomy technique.

Authors:  Emma Jones; Jonathan Clarke; Peng Tee Khaw
Journal:  Curr Opin Ophthalmol       Date:  2005-04       Impact factor: 3.761

2.  The mucosal barrier at implant abutments of different materials.

Authors:  Maria Welander; Ingemar Abrahamsson; Tord Berglundh
Journal:  Clin Oral Implants Res       Date:  2008-05-19       Impact factor: 5.977

3.  High-precision, cost-effective cutting system for producing thin sections of oral tissues containing dental implants.

Authors:  A Piattelli; A Scarano; M Quaranta
Journal:  Biomaterials       Date:  1997-04       Impact factor: 12.479

4.  Silicone elastomer implantation cyclodialysis. A negative report.

Authors:  G L Portney
Journal:  Arch Ophthalmol       Date:  1973-01

5.  Hydrostatic pressure of the suprachoroidal space.

Authors:  K Emi; J E Pederson; C B Toris
Journal:  Invest Ophthalmol Vis Sci       Date:  1989-02       Impact factor: 4.799

6.  Prevention of early postoperative hypotony by partial ligation of silicone tube in Ahmed glaucoma valve implantation.

Authors:  C Kee
Journal:  J Glaucoma       Date:  2001-12       Impact factor: 2.503

7.  In vivo analysis of conjunctiva in gold micro shunt implantation for glaucoma.

Authors:  Leonardo Mastropasqua; Luca Agnifili; Marco Ciancaglini; Mario Nubile; Paolo Carpineto; Vincenzo Fasanella; Michele Figus; Stefano Lazzeri; Marco Nardi
Journal:  Br J Ophthalmol       Date:  2010-08-30       Impact factor: 4.638

8.  Randomized evaluation of the trabecular micro-bypass stent with phacoemulsification in patients with glaucoma and cataract.

Authors:  Thomas W Samuelson; L Jay Katz; Jeffrey M Wells; Yi-Jing Duh; Jane Ellen Giamporcaro
Journal:  Ophthalmology       Date:  2010-09-15       Impact factor: 12.079

9.  Initial clinical experience with the Ahmed Glaucoma Valve implant.

Authors:  A L Coleman; R Hill; M R Wilson; N Choplin; R Kotas-Neumann; M Tam; J Bacharach; W C Panek
Journal:  Am J Ophthalmol       Date:  1995-07       Impact factor: 5.258

10.  Aqueous humor in glaucomatous eyes contains an increased level of TGF-beta 2.

Authors:  R C Tripathi; J Li; W F Chan; B J Tripathi
Journal:  Exp Eye Res       Date:  1994-12       Impact factor: 3.467

View more
  17 in total

1.  Safety and effectiveness of gold glaucoma micro shunt for reducing intraocular pressure in Japanese patients with open angle glaucoma.

Authors:  Masaki Tanito; Etsuo Chihara
Journal:  Jpn J Ophthalmol       Date:  2017-06-09       Impact factor: 2.447

2.  [STARflo-a suprachoroidal drainage implant in glaucoma surgery].

Authors:  S König; C W Hirneiß
Journal:  Ophthalmologe       Date:  2018-08       Impact factor: 1.059

3.  Development of an Experimental Drug Eluting Suprachoroidal Microstent as Glaucoma Drainage Device.

Authors:  Marina Hovakimyan; Stefan Siewert; Wolfram Schmidt; Katrin Sternberg; Thomas Reske; Oliver Stachs; Rudolf Guthoff; Andreas Wree; Martin Witt; Klaus-Peter Schmitz; Reto Allemann
Journal:  Transl Vis Sci Technol       Date:  2015-06-30       Impact factor: 3.283

4.  Clinical options for the reduction of elevated intraocular pressure.

Authors:  Laura Crawley; Sohaib M Zamir; Maria F Cordeiro; Li Guo
Journal:  Ophthalmol Eye Dis       Date:  2012-04-30

5.  Long-term outcome of supraciliary gold micro shunt in refractory glaucoma.

Authors:  Michele Figus; Pasquale Loiudice; Andrea Passani; Laura Perciballi; Luca Agnifili; Marco Nardi; Chiara Posarelli
Journal:  Acta Ophthalmol       Date:  2021-07-27       Impact factor: 3.988

Review 6.  Device-modified trabeculectomy for glaucoma.

Authors:  Xue Wang; Rabeea Khan; Anne Coleman
Journal:  Cochrane Database Syst Rev       Date:  2015-12-01

7.  Retrospective analysis of the success and safety of Gold Micro Shunt Implantation in glaucoma.

Authors:  Arno Hueber; Sigrid Roters; Jens F Jordan; Walter Konen
Journal:  BMC Ophthalmol       Date:  2013-07-18       Impact factor: 2.209

Review 8.  Update on Minimally Invasive Glaucoma Surgery (MIGS) and New Implants.

Authors:  Lívia M Brandão; Matthias C Grieshaber
Journal:  J Ophthalmol       Date:  2013-11-27       Impact factor: 1.909

9.  Clinical and morphological evaluation of gold micro shunt after unsuccessful surgical treatment of patients with primary open-angle glaucoma.

Authors:  M Rękas; B Pawlik; B Grala; W Kozłowski
Journal:  Eye (Lond)       Date:  2013-07-19       Impact factor: 3.775

10.  Subconjunctival bevacizumab to augment trabeculectomy with mitomycin C in the management of failed glaucoma surgery.

Authors:  Ahmed M Saeed; Tarek Tawfeek AboulNasr
Journal:  Clin Ophthalmol       Date:  2014-09-15
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.