| Literature DB >> 21859593 |
Huan Huang1, Ming Yang, Wangfu Zang, Shunjie Wu, Yafei Pang.
Abstract
Mock circulatory loops (MCLs) have been widely used to test left ventricular assist devices. The hydraulic properties of the mock systemic arterial system are usually described by two alternative four-element windkessel (W4) models. Compared with three-element windkessel model, their parameters, especially the inertial term, are much more difficult to estimate. In this paper, an estimator based on the iterated unscented Kalman filter (IUKF) algorithm is proposed to identify model parameters. Identifiability of these parameters for different measurements is described. Performance of the estimator for different model structures is first evaluated using numerical simulation data contaminated with artificial noise. An MCL is developed to test the proposed algorithm. Parameter estimates for different models are compared with the calculated values derived from the mechanical and hydraulic properties of the MCL to validate model structures. In conclusion, the W4 model with an inertance and an aortic characteristic resistance arranged in series is proposed to represent the mock systemic arterial system. Once model structure is appropriately selected, IUKF can provide reasonable estimation accuracy in a limited time and may be helpful for future clinical applications.Entities:
Mesh:
Year: 2011 PMID: 21859593 DOI: 10.1109/TBME.2011.2161477
Source DB: PubMed Journal: IEEE Trans Biomed Eng ISSN: 0018-9294 Impact factor: 4.538