BACKGROUND: New hardware and software algorithms in electromagnetic tracking for computer assisted surgery (CAS) have been developed. We aimed to compare electromagnetic tracking for navigated procedures in frontal skull base surgery to optical tracking.
METHODS: Target registration error (TRE) was determined in 6 anatomic specimens in an experimental wet-lab. As targets, 6 titanium screws were evenly distributed over the surgical areas of interest from the frontal sinus to the clivus. Optical tracking and electromagnetic tracking was evaluated in identical software environment using a last generation commercially available navigation system.
RESULTS: Submillimetric application accuracy could be achieved with both tracking modalities. Optical was more accurate than electromagnetic tracking and its reliability was better. Target position did not influence TRE, however TRE varied significantly from skull to skull.
CONCLUSIONS: Although less accurate than optical tracking, electromagnetic tracking still offers excellent accuracy and reliability for anterior skull base surgery. Electromagnetic tracking is not dependent on direct line of sight between its hardware components and therefore easily integrated even in cluttered operating theatres.

Mesh:

Year:  2011        PMID: 21858270     DOI: 10.4193/Rhino10.177

Source DB:  PubMed          Journal:  Rhinology        ISSN: 0300-0729            Impact factor:   3.681


  7 in total

Review 1.  Comprehensive review on endonasal endoscopic sinus surgery.

Authors:  Rainer K Weber; Werner Hosemann
Journal:  GMS Curr Top Otorhinolaryngol Head Neck Surg       Date:  2015-12-22

2.  The use of a head-mounted display in oral implantology: a feasibility study.

Authors:  Balázs Vigh; Steffen Müller; Oliver Ristow; Herbert Deppe; Stuart Holdstock; Jürgen den Hollander; Nassir Navab; Timm Steiner; Bettina Hohlweg-Majert
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-06-20       Impact factor: 2.924

3.  Distorter Characterisation Using Mutual Inductance in Electromagnetic Tracking.

Authors:  Herman Alexander Jaeger; Pádraig Cantillon-Murphy
Journal:  Sensors (Basel)       Date:  2018-09-12       Impact factor: 3.576

4.  Prospects and limitations of different registration modalities in electromagnetic ENT navigation.

Authors:  Eric Soteriou; Juergen Grauvogel; Roland Laszig; Tanja Daniela Grauvogel
Journal:  Eur Arch Otorhinolaryngol       Date:  2016-05-05       Impact factor: 2.503

5.  Missed paranasal sinus compartments in sinus surgery with and without image-guidance systems: a pilot feasibility study.

Authors:  Aris I Giotakis; Florian Kral; Wolfgang Freysinger; Stefan Markart; Herbert Riechelmann
Journal:  Int J Comput Assist Radiol Surg       Date:  2019-03-06       Impact factor: 2.924

6.  Intraoperative ultrasound-guided compared to stereotactic navigated ventriculoperitoneal shunt placement: study protocol for a randomised controlled study.

Authors:  Severina Leu; Florian Halbeisen; Luigi Mariani; Jehuda Soleman
Journal:  Trials       Date:  2021-05-19       Impact factor: 2.279

7.  Proof-of-concept of a laser mounted endoscope for touch-less navigated procedures.

Authors:  Florian Kral; Oezguer Gueler; Martina Perwoeg; Zoltan Bardosi; Elisabeth J Puschban; Herbert Riechelmann; Wolfgang Freysinger
Journal:  Lasers Surg Med       Date:  2013-06-04       Impact factor: 4.025

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.