Compound-specific chlorine isotope analysis: a comparison of gas chromatography/isotope ratio mass spectrometry and gas chromatography/quadrupole mass spectrometry methods in an interlaboratory study.

Anat Bernstein1, Orfan Shouakar-Stash, Karin Ebert, Christine Laskov, Daniel Hunkeler, Simon Jeannottat, Kaori Sakaguchi-Söder, Jens Laaks, Maik A Jochmann, Stefan Cretnik, Johannes Jager, Stefan B Haderlein, Torsten C Schmidt, Ramon Aravena, Martin Elsner.   

Abstract

Chlorine isotope analysis of chlorinated hydrocarbons like trichloroethylene (TCE) is of emerging demand because these species are important environmental pollutants. Continuous flow analysis of noncombusted TCE molecules, either by gas chromatography/isotope ratio mass spectrometry (GC/IRMS) or by GC/quadrupole mass spectrometry (GC/qMS), was recently brought forward as innovative analytical solution. Despite early implementations, a benchmark for routine applications has been missing. This study systematically compared the performance of GC/qMS versus GC/IRMS in six laboratories involving eight different instruments (GC/IRMS, Isoprime and Thermo MAT-253; GC/qMS, Agilent 5973N, two Agilent 5975C, two Thermo DSQII, and one Thermo DSQI). Calibrations of (37)Cl/(35)Cl instrument data against the international SMOC scale (Standard Mean Ocean Chloride) deviated between instruments and over time. Therefore, at least two calibration standards are required to obtain true differences between samples. Amount dependency of δ(37)Cl was pronounced for some instruments, but could be eliminated by corrections, or by adjusting amplitudes of standards and samples. Precision decreased in the order GC/IRMS (1σ ≈ 0.1‰), to GC/qMS (1σ ≈ 0.2-0.5‰ for Agilent GC/qMS and 1σ ≈ 0.2-0.9‰ for Thermo GC/qMS). Nonetheless, δ(37)Cl values between laboratories showed good agreement when the same external standards were used. These results lend confidence to the methods and may serve as a benchmark for future applications.
© 2011 American Chemical Society

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21851081     DOI: 10.1021/ac200516c

Source DB:  PubMed          Journal:  Anal Chem        ISSN: 0003-2700            Impact factor:   6.986


  2 in total

Review 1.  Photodegradation of pesticides using compound-specific isotope analysis (CSIA): a review.

Authors:  Guolu Cui; George Lartey-Young; Chong Chen; Limin Ma
Journal:  RSC Adv       Date:  2021-07-20       Impact factor: 4.036

2.  Modelling of C/Cl isotopic behaviour during chloroethene biotic reductive dechlorination: Capabilities and limitations of simplified and comprehensive models.

Authors:  Alice Badin; Fabian Braun; Landon J S Halloran; Julien Maillard; Daniel Hunkeler
Journal:  PLoS One       Date:  2018-08-22       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.