Accuracy of magnetic resonance imaging-directed frame-based stereotaxis.

Nova B Thani1, Arul Bala, Christopher R P Lind.   

Abstract

BACKGROUND: Accurate placement of a probe to the deep regions of the brain is an important part of neurosurgery. In the modern era, magnetic resonance image (MRI)-based target planning with frame-based stereotaxis is the most common technique.
OBJECTIVE: To quantify the inaccuracy in MRI-guided frame-based stereotaxis and to assess the relative contributions of frame movements and MRI distortion.
METHODS: The MRI-directed implantable guide-tube technique was used to place carbothane stylettes before implantation of the deep brain stimulation electrodes. The coordinates of target, dural entry point, and other brain landmarks were compared between preoperative and intraoperative MRIs to determine the inaccuracy.
RESULTS: The mean 3-dimensional inaccuracy of the stylette at the target was 1.8 mm (95% confidence interval [CI], 1.5-2.1. In deep brain stimulation surgery, the accuracy in the x and y (axial) planes is important; the mean axial inaccuracy was 1.4 mm (95% CI, 1.1-1.8). The maximal mean deviation of the head frame compared with brain over 24.1 ± 1.8 hours was 0.9 mm (95% CI, 0.5-1.1). The mean 3-dimensional inaccuracy of the dural entry point of the stylette was 1.8 mm (95% CI, 1.5-2.1), which is identical to that of the target.
CONCLUSION: Stylette positions did deviate from the plan, albeit by 1.4 mm in the axial plane and 1.8 mm in 3-dimensional space. There was no difference between the accuracies at the dura and the target approximately 70 mm deep in the brain, suggesting potential feasibility for accurate planning along the whole trajectory.

Entities:  

Mesh:

Year:  2012        PMID: 21849920     DOI: 10.1227/NEU.0b013e3182320bd6

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  4 in total

1.  Deep Brain Stimulation and Swimming Performance: A Randomized Within-Person Crossover Study.

Authors:  Sarah Katherine Morgan; Omar Khan Bangash; Nat Benjanuvatra; Megan Thorburn; Irné Du Plessis; Angela Jacques; Georgina Powers; Christopher R P Lind
Journal:  Neurol Clin Pract       Date:  2021-10

2.  The anatomy of the caudal zona incerta in rodents and primates.

Authors:  Charles Watson; Christopher R P Lind; Meghan G Thomas
Journal:  J Anat       Date:  2013-10-21       Impact factor: 2.610

3.  Intraoperative detection of blood vessels with an imaging needle during neurosurgery in humans.

Authors:  Hari Ramakonar; Bryden C Quirk; Rodney W Kirk; Jiawen Li; Angela Jacques; Christopher R P Lind; Robert A McLaughlin
Journal:  Sci Adv       Date:  2018-12-19       Impact factor: 14.136

4.  Image fusion pitfalls for cranial radiosurgery.

Authors:  Benjamin P Jonker
Journal:  Surg Neurol Int       Date:  2013-04-17
  4 in total