BACKGROUND: Insufflation of carbon dioxide (CO(2)) to the operative field has been used to prevent major organ injury attributed to air embolisms in cardiac surgery. However, it may be preferable to avoid hypercapnia induced by CO(2) insufflation, owing to its potentially harmful effect. To investigate the effectiveness of near-infrared spectroscopy (NIRS) as a possible method for continuous monitoring of arterial CO(2) tension during cardiac surgery, we evaluated the correlation between the change in arterial CO(2) tension and the change in regional cerebral oxygen saturation (rScO(2)) obtained from NIRS in as controlled a condition as possible.
METHODS: Thirty patients who underwent surgical correction for atrial or ventricular septal defects were enrolled in this study. Patients who had pulmonary hypertension or other intracardiac anomalies were excluded. Anesthetic and cardiopulmonary bypass (CPB) management were conducted according to our standard institutional practice. Data obtained from arterial blood gas analyses and corresponding regional cerebral oxygen saturation (rScO(2)) recorded from NIRS before and after the insufflations of CO(2) during CPB were used for analysis.
RESULTS: The change in arterial CO(2) tension correlated with the change in rScO(2) in the left hemisphere (r = 0.681, p <0.001, y = -1.393 + 0.547x) and right hemisphere (r = 0.690, p <0.001, y = -1.999 + 0.486x). To control the effects of other variables, including hematocrit and temperature, these relationship were not reduced (left hemisphere: r=0.678, p<0.001; right hemisphere: r=0.634, p<0.001).
CONCLUSIONS: Since the change in regional cerebral oxygen saturation was correlated with the change in arterial CO(2) tension during mild hypothermic CPB, NIRS might be a possible non-invasive method for monitoring of arterial CO(2) tension without incurring additional cost in this setting.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21844113     DOI: 10.1177/0267659111419034

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  3 in total

1.  A validation method for near-infrared spectroscopy based tissue oximeters for cerebral and somatic tissue oxygen saturation measurements.

Authors:  Paul B Benni; David MacLeod; Keita Ikeda; Hung-Mo Lin
Journal:  J Clin Monit Comput       Date:  2017-04-03       Impact factor: 2.502

2.  Effects of carbon dioxide insufflation on anastomosis remodeling at a carotid artery site in rabbits.

Authors:  Tuğra Gençpınar; Gökmen Akkaya; Çağatay Bilen; Pınar Akokay; Osman Yılmaz; Hudai Çatalyürek
Journal:  Kardiochir Torakochirurgia Pol       Date:  2018-09-24

3.  The effects of ventilation on left-to-right shunt and regional cerebral oxygen saturation: a self-controlled trial.

Authors:  Peiyi Li; Jun Zeng; Wei Wei; Jing Lin
Journal:  BMC Anesthesiol       Date:  2019-10-09       Impact factor: 2.217

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.