Literature DB >> 21843293

Results of animal experiments with the fourth model of the undulation pump total artificial heart.

Yusuke Abe1, Takashi Isoyama, Itsuro Saito, Wei Shi, Yusuke Inoue, Kohei Ishii, Hidemoto Nakagawa, Toshiya Ono, Minoru Ono, Kou Imachi.   

Abstract

Animal experiments using a total artificial heart in a goat are not easy to perform. The fourth model of the undulation pump total artificial heart (UPTAH4), which was designed to perform a long-term physiological experiment including pulsatile and nonpulsatile TAH operations with a conductance- and arterial pressure-based control method named 1/R control, was implanted in 31 goats weighing 38.5 to 60.4 kg (average of 46.8 kg). The 1/R control is a physiological flow control method of TAH developed with a conductance (1/R: reciprocal of a resistance) parallel circuit model. The survival periods were from 0.1 to 153 days (average of 14.5 days). The causes of termination were postoperative bleeding in eight goats, respiratory failure in five goats, device failure in 14 goats, dissected aneurysm in two goats, and thrombus in one goat. The thrombus case was the longest surviving goat. The respiratory failure tended to occur when the extracorporeal circulation time was prolonged. Autotransfusion was effective for the prolongation of survival time. The left-right balance control and the suction control were performed successfully in all goats. The 1/R control was performed for a long time in five goats that survived for more than 1 month. With three goats that survived for 48, 52, and 53 days mainly with the pulsatile mode, the 1/R control was stable. With a goat that survived for 73 days, the nonpulsatile mode with the 1/R control could be tested for 3 weeks. With the longest surviving goat that was maintained mainly with the pulsatile mode, the 1/R control was unstable, possibly due to the mismatching of the response time of the control system between the computer and the body. However, liver and kidney functions were almost normal, and the total protein level recovered. Further study to stabilize the 1/R control in the UPTAH is necessary.
© 2011, Copyright the Authors. Artificial Organs © 2011, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21843293     DOI: 10.1111/j.1525-1594.2011.01318.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  4 in total

1.  Appropriate control time constant in relation to characteristics of the baroreflex vascular system in 1/R control of the total artificial heart.

Authors:  Sora Mizuta; Itsuro Saito; Takashi Isoyama; Shintaro Hara; Terumi Yurimoto; Xinyang Li; Haruka Murakami; Toshiya Ono; Kunihiko Mabuchi; Yusuke Abe
Journal:  J Artif Organs       Date:  2017-05-17       Impact factor: 1.731

2.  The helical flow pump with a hydrodynamic levitation impeller.

Authors:  Yusuke Abe; Kohei Ishii; Takashi Isoyama; Itsuro Saito; Yusuke Inoue; Toshiya Ono; Hidemoto Nakagawa; Emiko Nakano; Kyoko Fukazawa; Kazuhiko Ishihara; Kazuyoshi Fukunaga; Minoru Ono; Kou Imachi
Journal:  J Artif Organs       Date:  2012-08-28       Impact factor: 1.731

Review 3.  An insight into short- and long-term mechanical circulatory support systems.

Authors:  Markus Ferrari; Peter Kruzliak; Kyriakos Spiliopoulos
Journal:  Clin Res Cardiol       Date:  2014-10-28       Impact factor: 5.460

Review 4.  The ongoing quest for the first total artificial heart as destination therapy.

Authors:  Annemijn Vis; Maziar Arfaee; Husain Khambati; Mark S Slaughter; Jan F Gummert; Johannes T B Overvelde; Jolanda Kluin
Journal:  Nat Rev Cardiol       Date:  2022-06-06       Impact factor: 32.419

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.