Assessment of the Stratos, a new pencil-beam bone densitometer: dosimetry, precision, and cross calibration.

Enrique Nalda1, Krishna Kunal Mahadea, Christophe Dematteï, Pierre-Olivier Kotzki, Jean-Pierre Pouget, Vincent Boudousq.   

Abstract

The goal of this study was to assess a new pencil-beam densitometer, the Stratos (Diagnostic Medical Systems, Pérols, France). Evaluation of the dosimetry and precision were done together with an in vivo cross-calibration study performed with the fan beam densitometer Discovery A (Hologic, Bedford, MA). The results indicated that the Stratos performed bone mineral density (BMD) measurements with a good precision, low radiation dose, and good agreement with the Discovery A. The air dose, measured by an ionization chamber, was 40 μGy. The effective dose was assessed using an anthropomorphic phantom and thermoluminescent detectors resulting in 1.96 and 0.31 μSv for a lumbar spine and proximal femur scan, respectively. The average scattered dose rate at a distance of 1m from the device was 1.06 and 1.21 μSv.h(-1) in the lumbar spine and left proximal femur scan mode, respectively. For the precision evaluation, 30 patients underwent 2 lumbar spine and 2 proximal femur scans with repositioning after each scan. The percentage root-mean-square coefficient of variation was 1.22%, 1.38%, 2.11%, and 0.86% for the lumbar spine (L1-L4), lumbar spine (L2-L4), femoral neck, and total hip, respectively. The cross-calibration studies were done on 57 patients (60 ± 9 yr). Lumbar spine, left neck, and left total hip mean BMD were 3.10% lower and 11.94% and 8.83% higher, respectively, with the Stratos compared with the Discovery A. Cross-calibration equations were calculated with a correlation coefficient of 98% (p<0.01) for the lumbar spine (L2-L4), 94% (p<0.01) for the left neck, and 92% (p<0.01) for the left total hip. After standardizing the Stratos measures using the cross-calibration equations, LIN's concordance correlation coefficient was 0.98, 0.93, and 0.92 for the lumbar spine (L2-L4), left neck, and total hip, respectively. Copyright Â
© 2011 The International Society for Clinical Densitometry. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21839660     DOI: 10.1016/j.jocd.2011.05.004

Source DB:  PubMed          Journal:  J Clin Densitom        ISSN: 1094-6950            Impact factor:   2.617


  2 in total

1.  Comparison of novel DXA system, Shimazu SONIALVISION G4, with GE-Lunar PRODIGY.

Authors:  Toki Takemoto; Takeshi Oguchi; Koji Oda
Journal:  J Bone Miner Metab       Date:  2021-06-11       Impact factor: 2.626

2.  Association between Anthropometric Indices, Body Composition and Bone Parameters in Thai Female Adolescents.

Authors:  Rapheeporn Khwanchuea; Chuchard Punsawad
Journal:  Indian J Pediatr       Date:  2017-08-11       Impact factor: 1.967

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.