Precision and accuracy of blood glucose measurements using three different instruments.

B Nowotny1, P J Nowotny, K Strassburger, M Roden.   

Abstract

AIMS: Assessment of insulin sensitivity by dynamic metabolic tests such as the hyperinsulinemic euglycemic clamp critically relies on the reproducible and fast measurement of blood glucose concentrations. Although various instruments have been developed over the last decades, little is known as to the accuracy and comparability. We therefore compared the best new instrument with the former gold standard instruments to measure glucose concentrations in metabolic tests.
METHODS: Fasting blood samples of 15 diabetic and 10 healthy subjects were collected into sodium-fluoride tubes, spiked with glucose (0, 2.8, 6.9 and 11.1 mmol/l) and measured either as whole blood (range 3.3-26.3 mmol/l) or following centrifugation as plasma (range 3.9-32.0 mmol/l). Plasma samples were analyzed in the YSI-2300 STAT plus (YSI), EKF Biosen C-Line (EKF) and the reference method, Beckman Glucose analyzer-II (BMG), whole blood samples in EKF instruments with YSI as reference method.
RESULTS: The average deviation of the EKF from the reference, BMG, was 3.0 ± 3.5% without any concentration-dependent variability. Glucose measurements by YSI were in good agreement with that by BMG (plasma) and EKF (plasma and whole blood) up to concentrations of 13.13 mmol/l (0.5 ± 3.7%), but deviation increased to -6.2 ± 3.8% at higher concentrations. Precision (n = 6) was ±2.2% (YSI), ±3.9% (EKF) and ±5.2% (BMG).
CONCLUSIONS: The EKF instrument is comparable regarding accuracy and precision to the reference method BMG and can be used in metabolic tests, while the YSI showed a systematic shift at higher glucose concentrations. Based on these results we decided to replace BMG with EKF instrument in metabolic tests.
© 2012 The Authors. Diabetic Medicine © 2012 Diabetes UK.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 21824188     DOI: 10.1111/j.1464-5491.2011.03406.x

Source DB:  PubMed          Journal:  Diabet Med        ISSN: 0742-3071            Impact factor:   4.359


  17 in total

1.  Analysis of "Accuracy evaluation of five blood glucose monitoring systems: the North American comparator trial".

Authors:  Paul A Fournier
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

2.  Fundamental Importance of Reference Glucose Analyzer Accuracy for Evaluating the Performance of Blood Glucose Monitoring Systems (BGMSs).

Authors:  Timothy S Bailey; Leslie J Klaff; Jane F Wallace; Carmine Greene; Scott Pardo; Bern Harrison; David A Simmons
Journal:  J Diabetes Sci Technol       Date:  2016-06-28

3.  Central and Peripheral Administration of Fibroblast Growth Factor 1 Improves Pancreatic Islet Insulin Secretion in Diabetic Mouse Models.

Authors:  Katherine G Tennant; Sarah R Lindsley; Melissa A Kirigiti; Cadence True; Paul Kievit
Journal:  Diabetes       Date:  2019-05-02       Impact factor: 9.461

4.  Estimates of insulin sensitivity from the intravenous-glucose-modified-clamp test depend on suppression of lipolysis in type 2 diabetes: a randomised controlled trial.

Authors:  Sabine Kahl; Bettina Nowotny; Simon Piepel; Peter J Nowotny; Klaus Strassburger; Christian Herder; Giovanni Pacini; Michael Roden
Journal:  Diabetologia       Date:  2014-07-22       Impact factor: 10.122

5.  A Comparison Study Between Point-of-Care Testing Systems and Central Laboratory for Determining Blood Glucose in Venous Blood.

Authors:  Huiping Wei; Fang Lan; Qitian He; Haiwei Li; Fuyong Zhang; Xue Qin; Shan Li
Journal:  J Clin Lab Anal       Date:  2016-08-25       Impact factor: 2.352

6.  Low-energy diets differing in fibre, red meat and coffee intake equally improve insulin sensitivity in type 2 diabetes: a randomised feasibility trial.

Authors:  Bettina Nowotny; Lejla Zahiragic; Alessandra Bierwagen; Stefan Kabisch; Jan B Groener; Peter J Nowotny; Ann Kristin Fleitmann; Christian Herder; Giovanni Pacini; Iris Erlund; Rikard Landberg; Hans-Ulrich Haering; Andreas F H Pfeiffer; Peter P Nawroth; Michael Roden
Journal:  Diabetologia       Date:  2014-11-26       Impact factor: 10.122

7.  Acute dietary fat intake initiates alterations in energy metabolism and insulin resistance.

Authors:  Elisa Álvarez Hernández; Sabine Kahl; Anett Seelig; Paul Begovatz; Martin Irmler; Yuliya Kupriyanova; Bettina Nowotny; Peter Nowotny; Christian Herder; Cristina Barosa; Filipa Carvalho; Jan Rozman; Susanne Neschen; John G Jones; Johannes Beckers; Martin Hrabě de Angelis; Michael Roden
Journal:  J Clin Invest       Date:  2017-01-23       Impact factor: 14.808

8.  Associations between central nervous system serotonin, fasting glucose, and hostility in African American females.

Authors:  Stephen H Boyle; Anastasia Georgiades; Beverly H Brummett; John C Barefoot; Ilene C Siegler; Wayne R Matson; Cynthia M Kuhn; Katherine Grichnik; Mark Stafford-Smith; Redford B Williams; Rima Kaddurah-Daouk; Richard S Surwit
Journal:  Ann Behav Med       Date:  2015-02

9.  Athletes feature greater rates of muscle glucose transport and glycogen synthesis during lipid infusion.

Authors:  Esther Phielix; Paul Begovatz; Sofiya Gancheva; Alessandra Bierwagen; Esther Kornips; Gert Schaart; Matthijs K C Hesselink; Patrick Schrauwen; Michael Roden
Journal:  JCI Insight       Date:  2019-11-01

10.  Mechanisms underlying the onset of oral lipid-induced skeletal muscle insulin resistance in humans.

Authors:  Bettina Nowotny; Lejla Zahiragic; Dorothea Krog; Peter J Nowotny; Christian Herder; Maren Carstensen; Toru Yoshimura; Julia Szendroedi; Esther Phielix; Peter Schadewaldt; Nanette C Schloot; Gerald I Shulman; Michael Roden
Journal:  Diabetes       Date:  2013-03-01       Impact factor: 9.461

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.