Ulrich Englich1, Peter Revesz, William Miller.
Abstract
In order to accurately measure the photon flux and to assist in aligning the beam, we have designed a modified beam stop device based on a photo diode integrated with the beam stop. The beam stop contains a small CdWO(4) crystal that completely stops the X-rays and at the same time produces photoluminescence proportional to the X-ray flux. The light is then guided to a photosensitive diode, using a flexible light pipe, to monitor the flux. With this device we achieve the goal of stopping the primary X-ray beam and simultaneously monitoring the X-ray intensity, thus eliminating the need for integrating ion-chambers into the capillary or collimator mount.Entities:
Year: 2011 PMID: 21822345 PMCID: PMC3150576 DOI: 10.1016/j.nima.2010.12.023