Literature DB >> 21817139

Mechanical thrombectomy in acute embolic stroke: preliminary results with the revive device.

Stefan Rohde1, Stefan Haehnel, Christian Herweh, Mirko Pham, Sibylle Stampfl, Peter A Ringleb, Martin Bendszus.   

Abstract

BACKGROUND AND
PURPOSE: The purpose of this study was to evaluate the safety and technical feasibility of a new thrombectomy device (Revive; Micrus Endovascular) in the endovascular treatment of acute ischemic stroke.
METHODS: Ten patients with acute large vessel occlusions were treated with the Revive device between October 2010 and December 2010. Mean National Institutes of Health Stroke Scale on admission was 19.0; mean duration of symptoms was 172 minutes. Recanalization was assessed using the Thrombolysis In Cerebral Infarction score. Clinical outcome (National Institutes of Health Stroke Scale) after thrombectomy was determined on Day 1, at discharge, and at Day 30.
RESULTS: Vessel recanalization (Thrombolysis In Cerebral Infarction 2b or 3) was successful in all patients without device-related complications. Mean National Institutes of Health Stroke Scale 24 hours after the intervention, at discharge, and at Day 30 was 14.0, 11.5, and 5.1, respectively. At Day 30, 6 patients had a clinical improvement of >8 points or an National Institutes of Health Stroke Scale of 0 to 1, 1 patient showed minor improvement, and 3 patients had died. Symptomatic intracranial hemorrhage occurred in 2 patients, of which 1 was fatal.
CONCLUSIONS: Thrombectomy with the Revive device in patients with stroke with acute large vessel occlusions demonstrated to be technically safe and highly effective. Clinical safety and efficacy have to be established in larger clinical trials.

Entities:  

Mesh:

Year:  2011        PMID: 21817139     DOI: 10.1161/STROKEAHA.111.616763

Source DB:  PubMed          Journal:  Stroke        ISSN: 0039-2499            Impact factor:   7.914


  24 in total

Review 1.  Safety and Efficacy of Mechanical Thrombectomy Using Stent Retrievers in the Endovascular Treatment of Acute Ischaemic Stroke: A Systematic Review.

Authors:  Janet Puñal-Riobóo; Gerardo Atienza; Miguel Blanco
Journal:  Interv Neurol       Date:  2015-07

Review 2.  Endovascular therapy in hyperacute ischaemic stroke: history and current status.

Authors:  Alex M Mortimer; Marcus D Bradley; Shelley A Renowden
Journal:  Interv Neuroradiol       Date:  2013-12-18       Impact factor: 1.610

3.  Mechanical thrombectomy with stent retrievers in acute basilar artery occlusion.

Authors:  M Möhlenbruch; S Stampfl; L Behrens; C Herweh; S Rohde; M Bendszus; C Hametner; S Nagel; P A Ringleb; M Pham
Journal:  AJNR Am J Neuroradiol       Date:  2013-11-28       Impact factor: 3.825

Review 4.  Neurothrombectomy in the treatment of acute ischaemic stroke.

Authors:  Olav Jansen; Axel Rohr
Journal:  Nat Rev Neurol       Date:  2013-10-15       Impact factor: 42.937

5.  Safety and efficacy of the Aperio thrombectomy device when compared to the Solitaire AB/FR and the Revive devices in a pulsatile flow system.

Authors:  Mahdi Saleh; John Nathan Spence; Sanjeev Nayak; Gillian Pearce; Christopher Tennuci; Christine Roffe
Journal:  Am J Cardiovasc Dis       Date:  2012-10-25

6.  Future trials of endovascular mechanical recanalisation therapy in acute ischemic stroke patients: a position paper endorsed by ESMINT and ESNR : part I: Current situation and major research questions.

Authors:  Jens Fiehler; Michael Söderman; Francis Turjman; Philip M White; Søren Jacob Bakke; Salvatore Mangiafico; Rüdiger von Kummer; Mario Muto; Christophe Cognard; Jan Gralla
Journal:  Neuroradiology       Date:  2012-09-05       Impact factor: 2.804

7.  [Industry-funded therapy studies: what is in the pipeline?].

Authors:  P D Schellinger; M Köhrmann; J Röther
Journal:  Nervenarzt       Date:  2012-10       Impact factor: 1.214

8.  Experimental evaluation of immediate recanalization effect and recanalization efficacy of a new thrombus retriever for acute stroke treatment in vivo.

Authors:  P Mordasini; C Brekenfeld; J V Byrne; U Fischer; M Arnold; S Jung; G Schroth; J Gralla
Journal:  AJNR Am J Neuroradiol       Date:  2012-07-26       Impact factor: 3.825

Review 9.  What is meant by "TICI"?

Authors:  J E Fugate; A M Klunder; D F Kallmes
Journal:  AJNR Am J Neuroradiol       Date:  2013-04-11       Impact factor: 3.825

10.  Recanalization with stent-retriever devices in patients with wake-up stroke.

Authors:  S Stampfl; P A Ringleb; S Haehnel; A Rocco; C Herweh; C Hametner; M Pham; M Moehlenbruch; M Bendszus; S Rohde
Journal:  AJNR Am J Neuroradiol       Date:  2012-12-20       Impact factor: 3.825