| Literature DB >> 21816600 |
Tae Yun Yeo1, Ji Suk Choi, Byung Kook Lee, Beob Soo Kim, Hwa In Yoon, Hyeong Yun Lee, Yong Woo Cho.
Abstract
Even low concentrations of endotoxins can be life-threatening. As such, continuous effort has been directed toward the development of sensitive and specific endotoxin detection systems. In this paper, we report the design and fabrication of a new electrochemical endotoxin sensor based on a human recombinant toll-like receptor 4 (rhTLR4) and myeloid differentiation-2 (MD-2) complex. The rhTLR4/MD-2 complex, which specifically binds to endotoxin, was immobilized on gold electrodes through a self-assembled monolayer (SAM) technique involving the use of dithiobis(succinimidyl undecanoate) (DSU). The surface topography of the electrodes at each fabrication stage was characterized with a nanosurface profiler and atomic force microscope (AFM). The electrochemical signals generated from interactions between the rhTLR4/MD-2 complex and the endotoxin were characterized by cyclic voltammetry (CV) and differential pulse voltammetry (DPV). A linear relationship between the peak current and endotoxin concentration was obtained in the range of 0.0005 to 5 EU/mL with a correlation coefficient (R(2)) of 0.978. The estimated limit of detection (LOD) was fairly low, 0.0002 EU/mL. The rhTLR4/MD-2 based sensors exhibited no current responses to dipalmitoylphosphatidylcholine (DPPC) bearing two lipid chains, which is structurally similar to endotoxin, indicating the high specificity of the sensors to endotoxin.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21816600 DOI: 10.1016/j.bios.2011.07.010
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618