Literature DB >> 21815378

An innovative iterative thresholding algorithm for tumour segmentation and volumetric quantification on SPECT images: Monte Carlo-based methodology and validation.

M Pacilio1, C Basile, S Shcherbinin, F Caselli, G Ventroni, D Aragno, L Mango, E Santini.   

Abstract

PURPOSE: Positron emission tomography (PET) and single-photon emission computed tomography (SPECT) imaging play an important role in the segmentation of functioning parts of organs or tumours, but an accurate and reproducible delineation is still a challenging task. In this work, an innovative iterative thresholding method for tumour segmentation has been proposed and implemented for a SPECT system. This method, which is based on experimental threshold-volume calibrations, implements also the recovery coefficients (RC) of the imaging system, so it has been called recovering iterative thresholding method (RIThM). The possibility to employ Monte Carlo (MC) simulations for system calibration was also investigated.
METHODS: The RIThM is an iterative algorithm coded using MATLAB: after an initial rough estimate of the volume of interest, the following calculations are repeated: (i) the corresponding source-to-background ratio (SBR) is measured and corrected by means of the RC curve; (ii) the threshold corresponding to the amended SBR value and the volume estimate is then found using threshold-volume data; (iii) new volume estimate is obtained by image thresholding. The process goes on until convergence. The RIThM was implemented for an Infinia Hawkeye 4 (GE Healthcare) SPECT/CT system, using a Jaszczak phantom and several test objects. Two MC codes were tested to simulate the calibration images: SIMIND and SimSet. For validation, test images consisting of hot spheres and some anatomical structures of the Zubal head phantom were simulated with SIMIND code. Additional test objects (flasks and vials) were also imaged experimentally. Finally, the RIThM was applied to evaluate three cases of brain metastases and two cases of high grade gliomas.
RESULTS: Comparing experimental thresholds and those obtained by MC simulations, a maximum difference of about 4% was found, within the errors (+/- 2% and +/- 5%, for volumes > or = 5 ml or < 5 ml, respectively). Also for the RC data, the comparison showed differences (up to 8%) within the assigned error (+/- 6%). ANOVA test demonstrated that the calibration results (in terms of thresholds or RCs at various volumes) obtained by MC simulations were indistinguishable from those obtained experimentally. The accuracy in volume determination for the simulated hot spheres was between -9% and 15% in the range 4-270 ml, whereas for volumes less than 4 ml (in the range 1-3 ml) the difference increased abruptly reaching values greater than 100%. For the Zubal head phantom, errors ranged between 9% and 18%. For the experimental test images, the accuracy level was within +/- 10%, for volumes in the range 20-110 ml. The preliminary test of application on patients evidenced the suitability of the method in a clinical setting.
CONCLUSIONS: The MC-guided delineation of tumor volume may reduce the acquisition time required for the experimental calibration. Analysis of images of several simulated and experimental test objects, Zubal head phantom and clinical cases demonstrated the robustness, suitability, accuracy, and speed of the proposed method. Nevertheless, studies concerning tumors of irregular shape and/or nonuniform distribution of the background activity are still in progress.

Entities:  

Mesh:

Year:  2011        PMID: 21815378     DOI: 10.1118/1.3590359

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Simplifying volumes-of-interest (VOIs) definition in quantitative SPECT: Beyond manual definition of 3D whole-organ VOIs.

Authors:  Esther M Vicente; Martin A Lodge; Steven P Rowe; Richard L Wahl; Eric C Frey
Journal:  Med Phys       Date:  2017-03-28       Impact factor: 4.071

2.  SPECT image segmentation for estimation of tumour volume and activity concentration in 177Lu-DOTATATE radionuclide therapy.

Authors:  Johan Gustafsson; Anna Sundlöv; Katarina Sjögreen Gleisner
Journal:  EJNMMI Res       Date:  2017-02-23       Impact factor: 3.138

3.  New strategy for automatic tumor segmentation by adaptive thresholding on PET/CT images.

Authors:  Mazen Moussallem; Pierre-Jean Valette; Alexandra Traverse-Glehen; Claire Houzard; Christophe Jegou; Francesco Giammarile
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.