New approaches to region of interest computed tomography.

Clemens Maass1, Michael Knaup, Marc Kachelriess.   

Abstract

PURPOSE: In classical x-ray CT, the diameter of the field of measurement (FOM) must not fall below the transversal diameter of the patient or specimen. Thereby, the ratio of the diameter of FOM and the number of transversal detector elements typically defines the spatial resolution. The authors aim at improving the spatial resolution within a region of interest (ROI) by a factor of 10-100 while maintaining artifact-free CT image reconstruction inside and outside the ROI. Two novel methods are proposed for artifact-free reconstruction of the truncated ROI scan (data weighting method and data filtering method) and compared with the gold standard (data completion method) for this problem.
METHODS: First, an overview scan with low spatial resolution and a large FOM that exceeds the object transversally is performed. Second, a high-resolution scan is performed, where the scanner's magnification is changed such that the FOM matches the ROI at the cost of laterally truncated projection data. The gold standard is forward projecting the low-resolution scan on the rays missing in the high-resolution scan. The authors propose the data filtering method, which uses the low-resolution reconstruction and calculates a high frequency correction term from the high-resolution scan, and the data weighting method, which reconstructs the truncated high-resolution data and calculates a detruncation image from the low-resolution data.
RESULTS: The methods are compared using a simulation of the Forbild head phantom and a measurement of a spinal disk implant. The results of the data weighting method and the data completion method show the same image quality. The data filtering method yields slightly inferior image quality that may still be sufficient for many applications. Both new methods considerably outperform the data completion method regarding the computational load.
CONCLUSIONS: The new ROI reconstruction methods are superior to the gold standard regarding the computational load. Comparing the image quality with the gold standard, the data filtering method is slightly inferior and the data weighting method yields equal quality.

Entities:  

Mesh:

Year:  2011        PMID: 21815362     DOI: 10.1118/1.3583696

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Time-resolved cardiac interventional cone-beam CT reconstruction from fully truncated projections using the prior image constrained compressed sensing (PICCS) algorithm.

Authors:  Pascal Thériault Lauzier; Jie Tang; Guang-Hong Chen
Journal:  Phys Med Biol       Date:  2012-04-05       Impact factor: 3.609

2.  Scout-view assisted interior micro-CT.

Authors:  Kriti Sen Sharma; Christian Holzner; Dragoş M Vasilescu; Xin Jin; Shree Narayanan; Masoud Agah; Eric A Hoffman; Hengyong Yu; Ge Wang
Journal:  Phys Med Biol       Date:  2013-06-04       Impact factor: 3.609

3.  Augmented laminography, a correlative 3D imaging method for revealing the inner structure of compressed fossils.

Authors:  Marcus Zuber; Michael Laaß; Elias Hamann; Sophie Kretschmer; Norbert Hauschke; Thomas van de Kamp; Tilo Baumbach; Thomas Koenig
Journal:  Sci Rep       Date:  2017-01-27       Impact factor: 4.379

4.  Clinical Micro-CT Empowered by Interior Tomography, Robotic Scanning, and Deep Learning.

Authors:  Mengzhou Li; Zheng Fang; Wenxiang Cong; Chuang Niu; Weiwen Wu; Josef Uher; James Bennett; Jay T Rubinstein; G E Wang
Journal:  IEEE Access       Date:  2020-12-21       Impact factor: 3.367

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.